Vergleichsstudie in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2025. Verfasst von Rochelle E Wong und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of Gastroenterology and Hepatology, University of Southern California Keck Medical Center, 1510 San Pablo St., HC1 - Suite 200, Los Angeles,…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Liver, Adipose Tissue, Fatty Liver. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science Foundation, University of Southern California.
Abstract (Originalsprache)
Background
Proton density fat fraction (PDFF)— the ratio of unconfounded fat signal to the sum of the unconfounded fat and water signals, is a valuable quantitative imaging biomarker of metabolic associated steatotic liver disease (MASLD) widely applied in clinical practice and clinical trials. PDFF of the liver is commonly measured using 3 T MRI systems. However, low-field systems are increasingly favored due to lower cost, improved safety profile, minimized artifacts around metallic implants, and enhanced patient comfort.
Objective
In this pilot study, we used knowledge of standardized and widely used 3 T liver PDFF protocols, and adapted parameters to be appropriate for the 0.55 T MRI. We evaluate a liver fat quantification protocol at 0.55 T compared to a standard clinical 3 T protocol to measure liver fat in patients with MASLD. Material and methods Eight adult patients (average age 53.6 ± 13.6 years, 5 females) with ≥ 5% PDFF on 3 T MRI underwent a 0.55 T MRI PDFF protocol within 90 days. To keep the acquisition time to be within a reasonable breath hold duration and with reasonable signal-to-noise ratio (SNR), four echoes were acquired at a lower resolution and fewer number of slices at 0.55 T compared to 3 T which uses a 6-echo multi-echo Dixon volumetric interpolated breath hold examination (VIBE) protocol. PDFF quantification accuracy of the 0.55 T approach was evaluated using a commercial PDFF phantom and in vivo.
Results
In the phantom, there was excellent match ( R 2 > 0.999) between PDFF estimated by 0.55 T MRI and ground truth. Mean in vivo 3 T MRI-PDFF was 16.5%, compared to 16.3% 0.55 T MRI-PDFF (correlation coefficient r = 0.99). The Bland–Altman analysis showed good agreement of in vivo PDFF measurements across 0.55 T and 3 T estimating a bias or mean difference of − 0.25% and the limits of agreements (LoA) of − 3.98% and 3.48%.
Discussion
Our data demonstrate that 0.55 T MRI is feasible and comparable to 3 T MRI in quantifying liver PDFF among patients with MASLD.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Gastroenterology and Hepatology, University of Southern California Keck Medical Center, 1510 San Pablo St., HC1 - Suite 200, Los Angeles, CA, 90033, USA. Rochelle.wong@med.usc.edu.
- Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, CA, USA.
- Department of Radiological Sciences, University of California Los Angeles, Los Angeles, CA, USA.
- Siemens Medical Solutions, Los Angeles, CA, USA.
- Division of Gastroenterology and Hepatology, University of Southern California Keck Medical Center, 1510 San Pablo St., HC1 - Suite 200, Los Angeles, CA, 90033, USA.
Förderung
- National Science Foundation
- University of Southern California
Zitieren
APA
Wong, R. E., Tasdelen, B., Tian, Y., Hwang, D., Cui, S. X., Yuan, L., & Nayak, K. S. (2025). In-vivo liver proton density fat fraction quantification at 0.55 T: a pilot study with comparison against 3 T MRI. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-025-01277-9
Vancouver
Wong RE, Tasdelen B, Tian Y, Hwang D, Cui SX, Yuan L, et al. In-vivo liver proton density fat fraction quantification at 0.55 T: a pilot study with comparison against 3 T MRI. Magnetic Resonance Materials in Physics, Biology and Medicine. 2025. doi:10.1007/s10334-025-01277-9
BibTeX
@article{Wong2025,
author = {Wong, Rochelle E and Tasdelen, Bilal and Tian, Ye and Hwang, Darryl and Cui, Sophia X and Yuan, Liyun and Nayak, Krishna S},
title = {In-vivo liver proton density fat fraction quantification at 0.55 T: a pilot study with comparison against 3 T MRI},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2025},
doi = {10.1007/s10334-025-01277-9}
}
EUVIMED