Radiofrequency ‐transparent local B 0 shimming coils using float traps

Changzhe Liu, Hao Liang, Ming Lu, John C Gore, Saikat Sengupta iD, Xinqiang Yan iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30361 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Changzhe Liu und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, Vanderbilt University, Nashville, Tennessee, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Static field (B 0 ) inhomogeneities present a major challenge in high‐field MRI. Multicoil shimming using independent, local, direct‐current (DC) shim coils has emerged as a powerful and flexible technique to address this issue. However, many‐turn DC coils can lead to significant mutual coupling with radiofrequency (RF) coils, causing transmit field (B 1 + ) distortions and signal‐to‐noise ratio degradation.

Methods

We introduce an innovative RF‐transparent DC coil that performs B 0 shimming while minimizing RF performance impact. The design incorporates float traps to maintain high RF impedance, allowing flexible placement relative to the RF coil without compromising signal‐to‐noise ratio or affecting B 1 + . We fabricated square‐shaped DC coils with float traps for 3T MRI and compared them with conventional DC coils. To demonstrate high ΔB 0 /Amp efficiency, we conducted a B 0 shimming experiment around a metal hip implant.

Results

Bench tests and MRI experimental results demonstrated that the RF‐transparent DC coil effectively minimized RF interference, preserved signal‐to‐noise ratio, and maintained B 1 + , even when placed near the RF receive coil. Additionally, the DC coil significantly improved B 0 homogeneity near metal implants and substantially reduced image distortion.

Conclusion

The RF‐transparent DC coil offers a flexible, effective solution for managing B 0 inhomogeneities, paving the way for integrating multiturn DC coils in clinical MRI settings without extensive hardware modifications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingSignal-To-Noise RatioRadio WavesEquipment DesignPhantoms, ImagingHumansImage Processing, Computer-Assisted

Stichwörter der Autor:innen

B0 shimmingRF transparencyimplantmulticoil shimmingsignal‐to‐noise ratio (SNR)transmit field

Autor:innen und Institutionen

  • Department of Electrical and Computer Engineering, Vanderbilt University, Nashville, Tennessee, USA.
  • Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee, USA.

ORCID-Profile: Saikat Sengupta, Xinqiang Yan

Förderung

  • NIBIB NIH HHS
  • NIH HHS

Zitieren

APA

Liu, C., Liang, H., Lu, M., Gore, J. C., Sengupta, S., & Yan, X. (2025). Radiofrequency ‐transparent local B 0 shimming coils using float traps. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30361

Vancouver

Liu C, Liang H, Lu M, Gore JC, Sengupta S, Yan X. Radiofrequency ‐transparent local B 0 shimming coils using float traps. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30361

BibTeX
@article{Liu2025,
  author = {Liu, Changzhe and Liang, Hao and Lu, Ming and Gore, John C and Sengupta, Saikat and Yan, Xinqiang},
  title = {Radiofrequency ‐transparent local B 0 shimming coils using float traps},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30361}
}
Kostenlos starten →