Physics‐guided self‐supervised learning: Demonstration for generalized RF pulse design
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30307 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2025. Written by Albert Jang and 2 co-authors. Institution: Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA. Major MEDLINE topics: Magnetic Resonance Imaging, Algorithms, Radio Waves. Cited 9 times so far (Crossref); 42 references. The full text is freely available (open access, CC BY 4.0). Funded by: NIBIB NIH HHS, NIAMS NIH HHS, National Institute of Arthritis and Musculoskeletal and Skin Diseases.
Abstract (original language)
Purpose
To introduce a new method for generalized RF pulse design using physics‐guided self‐supervised learning (GPS), which uses the Bloch equations as the guiding physics model. Theory and Methods The GPS framework consists of a neural network module and a physics module, where the physics module is a Bloch simulator for MRI applications. For RF pulse design, the neural network module maps an input target profile to an RF pulse, which is subsequently loaded into the physics module. Through the supervision of the physics module, the neural network module designs an RF pulse corresponding to the target profile. GPS was applied to design 1D selective, ‐insensitive, saturation, and multidimensional RF pulses, each conventionally requiring dedicated design algorithms. We further demonstrate our method's flexibility and versatility by compensating for experimental and scanner imperfections through online adaptation.
Results
Both simulations and experiments show that GPS can design a variety of RF pulses with corresponding profiles that agree well with the target input. Despite these verifications, GPS‐designed pulses have unique differences compared to conventional designs, such as achieving ‐insensitivity using different mechanisms and using non‐sampled regions of the conventional design to lower its peak power. Experiments, both ex vivo and in vivo, further verify that it can also be used for online adaptation to correct system imperfections, such as / inhomogeneity.
Conclusion
This work demonstrates the generalizability, versatility, and flexibility of the GPS method for designing RF pulses and showcases its utility in several applications.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
- Harvard Medical School, Boston, Massachusetts, USA.
ORCID profiles: Albert Jang, Fang Liu
Funding
- NIBIB NIH HHS
- NIAMS NIH HHS
- National Institute of Arthritis and Musculoskeletal and Skin Diseases
- National Institute of Biomedical Imaging and Bioengineering
Cite
APA
Jang, A., He, X., & Liu, F. (2025). Physics‐guided self‐supervised learning: Demonstration for generalized RF pulse design. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30307
Vancouver
Jang A, He X, Liu F. Physics‐guided self‐supervised learning: Demonstration for generalized RF pulse design. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30307
BibTeX
@article{Jang2025,
author = {Jang, Albert and He, Xingxin and Liu, Fang},
title = {Physics‐guided self‐supervised learning: Demonstration for generalized RF pulse design},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30307}
}
EUVIMED