SPICER: Self‐supervised learning for MRI with automatic coil sensitivity estimation and reconstruction

Yuyang Hu iD, Weijie Gan iD, Chunwei Ying iD, Tongyao Wang, Cihat Eldeniz iD, Jiaming Liu, Yasheng Chen, Hongyu An iD, Ulugbek S Kamilov iD

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30121 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Yuyang Hu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Systems Engineering, Washington University in St. Louis, St. Louis, Missouri. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Neural Networks, Computer, Algorithms. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 56 Literaturangaben. Gefördert von: NIBIB NIH HHS, NIH HHS, National Science Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To introduce a novel deep model‐based architecture (DMBA), SPICER, that uses pairs of noisy and undersampled k‐space measurements of the same object to jointly train a model for MRI reconstruction and automatic coil sensitivity estimation.

Methods

SPICER consists of two modules to simultaneously reconstructs accurate MR images and estimates high‐quality coil sensitivity maps (CSMs). The first module, CSM estimation module, uses a convolutional neural network (CNN) to estimate CSMs from the raw measurements. The second module, DMBA‐based MRI reconstruction module, forms reconstructed images from the input measurements and the estimated CSMs using both the physical measurement model and learned CNN prior. With the benefit of our self‐supervised learning strategy, SPICER can be efficiently trained without any fully sampled reference data.

Results

We validate SPICER on both open‐access datasets and experimentally collected data, showing that it can achieve state‐of‐the‐art performance in highly accelerated data acquisition settings (up to ). Our results also highlight the importance of different modules of SPICER—including the DMBA, the CSM estimation, and the SPICER training loss—on the final performance of the method. Moreover, SPICER can estimate better CSMs than pre‐estimation methods especially when the ACS data is limited.

Conclusion

Despite being trained on noisy undersampled data, SPICER can reconstruct high‐quality images and CSMs in highly undersampled settings, which outperforms other self‐supervised learning methods and matches the performance of the well‐known E2E‐VarNet trained on fully sampled ground‐truth data.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage Processing, Computer-AssistedNeural Networks, ComputerAlgorithmsHumansSupervised Machine LearningBrainDeep LearningPhantoms, Imaging

Stichwörter der Autor:innen

coil sensitivity estimationdeep learningimage reconstructioninverse problemsparallel MRI

Autor:innen und Institutionen

  • Department of Electrical and Systems Engineering, Washington University in St. Louis, St. Louis, Missouri.
  • Department of Computer Science and Engineering, Washington University in St. Louis, St. Louis, Missouri.
  • Mallinckrodt Institute of Radiology, Washington University in St. Louis, St. Louis, Missouri.
  • Department of Biomedical Engineering, Washington University in St. Louis, St. Louis, Missouri.
  • Department of Neurology, Washington University in St. Louis, St. Louis, Missouri.

ORCID-Profile: Yuyang Hu, Weijie Gan, Chunwei Ying, Cihat Eldeniz, Hongyu An, Ulugbek S Kamilov

Förderung

  • NIBIB NIH HHS
  • NIH HHS
  • National Science Foundation
  • National Institutes of Health

Zitieren

APA

Hu, Y., Gan, W., Ying, C., Wang, T., Eldeniz, C., Liu, J., Chen, Y., An, H., & Kamilov, U. S. (2024). SPICER: Self‐supervised learning for MRI with automatic coil sensitivity estimation and reconstruction. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30121

Vancouver

Hu Y, Gan W, Ying C, Wang T, Eldeniz C, Liu J, et al. SPICER: Self‐supervised learning for MRI with automatic coil sensitivity estimation and reconstruction. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30121

BibTeX
@article{Hu2024,
  author = {Hu, Yuyang and Gan, Weijie and Ying, Chunwei and Wang, Tongyao and Eldeniz, Cihat and Liu, Jiaming and Chen, Yasheng and An, Hongyu and Kamilov, Ulugbek S},
  title = {SPICER: Self‐supervised learning for MRI with automatic coil sensitivity estimation and reconstruction},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30121}
}
Kostenlos starten →