Local and whole‐body SAR in UHF body imaging: Implications for SAR matrix compression
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30306 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Thomas M Fiedler und 2 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Whole Body Imaging, Data Compression. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Purpose
Transmit arrays for body imaging have characteristics of both volume and local transmit coils. This study evaluates two specific absorption rate (SAR) aspects, local and whole‐body SAR, of arrays for body imaging at 7 T and also for a 3 T birdcage.
Methods
Simulations were performed for six antenna arrays at 7 T and one 3 T birdcage. Local SAR matrices and the whole‐body SAR matrix were computed and evaluated with random shims. A set of reduced local SAR matrices was determined by removing all matrices dominated by the whole‐body SAR matrix.
Results
The results indicate that all RF transmit coils for body imaging in this study are constrained by the local SAR limit. The ratio between local and whole‐body SAR is nevertheless smaller for arrays with large FOV, as these arrays also expose a larger part of the human body. By using the whole‐body SAR matrix, the number of local SAR matrices can be reduced (e.g., 33.3% matrices remained for an 8‐channel local array and 89.7% for a 30‐channel remote array; 12.1% for the 3 T birdcage).
Conclusion
For transmit antenna arrays used for body imaging at 7 T as well as for the 3 T birdcage, all evaluated cases show that the local SAR limit was reached before reaching the whole‐body SAR limit. Nevertheless, the whole‐body SAR matrix can be used to reduce the number of local SAR matrices, which is important to reduce memory and computing time for a virtual observation points (VOP) compression. This step can be included as a pre‐compression prior to a VOP compression.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
- Erwin L. Hahn Institute for MRI, University Duisburg-Essen, Essen, Germany.
- Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany.
- Faculty of Medicine, Heidelberg University, Heidelberg, Germany.
ORCID-Profile: Thomas M Fiedler, Stephan Orzada
Zitieren
APA
Fiedler, T. M., Ladd, M. E., & Orzada, S. (2025). Local and whole‐body SAR in UHF body imaging: Implications for SAR matrix compression. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30306
Vancouver
Fiedler TM, Ladd ME, Orzada S. Local and whole‐body SAR in UHF body imaging: Implications for SAR matrix compression. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30306
BibTeX
@article{Fiedler2025,
author = {Fiedler, Thomas M and Ladd, Mark E and Orzada, Stephan},
title = {Local and whole‐body SAR in UHF body imaging: Implications for SAR matrix compression},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30306}
}
EUVIMED