Accelerated B1+$$ {\mathrm{B}}_1^{+} $$ mapping and robust parallel transmit pulse design for heart and prostate imaging at 7 T
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30185 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2024. Written by Nico Egger and 11 co-authors. Institution: Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany. Major MEDLINE topics: Prostate, Heart, Magnetic Resonance Imaging, Algorithms. Cited 6 times so far (Crossref); 46 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Purpose
To investigate the impact of reduced k‐space sampling on mapping and the resulting impact on phase shimming and dynamic/universal parallel transmit (pTx) RF pulse design.
Methods
Channel‐wise 3D maps were measured at 7 T in 35 and 23 healthy subjects for the heart and prostate region, respectively. With these maps, universal phase shims optimizing homogeneity and efficiency were designed for heart and prostate imaging. In addition, universal 4kT‐point pulses were designed for the heart. Subsequently, individual phase shims and individual 4kT‐pulses were designed based on maps with different acceleration factors and tested on the original maps. The performance of the pulses was compared by evaluating their coefficients of variation (CoV), efficiencies and specific energy doses (SED). Furthermore, validation measurements were carried out for one heart and one prostate subject.
Results
For both organs, the universal phase shims showed significantly higher efficiencies and lower CoVs compared to the vendor provided default shim, but could still be improved with individual phase shims based on accelerated maps (acquisition time = 30 s). In the heart, the universal 4kT‐pulse achieved significantly lower CoVs than tailored phase shims. Tailored 4kT‐pulses based on accelerated maps resulted in even further reduced CoVs or a 2.5‐fold reduction in SED at the same CoVs as the universal 4kT‐pulse.
Conclusion
Accelerated maps can be used for the design of tailored pTx pulses for prostate and cardiac imaging at 7 T, which further improve homogeneity, efficiency, or SED compared to universal pulses.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
- Electrical Engineering and Information Technology, University of Applied Sciences - FH Aachen, Aachen, Germany.
- Siemens Healthcare, Erlangen, Bavaria, Germany.
- Rapid Biomedical, Rimpar, Germany.
- Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.
- Division of Medical Physics in Radiology, German Cancer Research Centre (DKFZ), Heidelberg, Germany.
ORCID profiles: Nico Egger, Laurent Ruck, Jürgen Herrler, Sebastian Schmitter, Armin Michael Nagel
Cite
APA
Egger, N., Nagelstraßer, S., Wildenberg, S., Bitz, A., Ruck, L., Herrler, J., Meixner, C. R., Kimmlingen, R., Lanz, T., Schmitter, S., Uder, M., & Nagel, A. M. (2024). Accelerated B1+$$ {\mathrm{B}}_1^{+} $$ mapping and robust parallel transmit pulse design for heart and prostate imaging at 7 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30185
Vancouver
Egger N, Nagelstraßer S, Wildenberg S, Bitz A, Ruck L, Herrler J, et al. Accelerated B1+$$ {\mathrm{B}}_1^{+} $$ mapping and robust parallel transmit pulse design for heart and prostate imaging at 7 T. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30185
BibTeX
@article{Egger2024,
author = {Egger, Nico and Nagelstraßer, Sophia and Wildenberg, Saskia and Bitz, Andreas and Ruck, Laurent and Herrler, Jürgen and Meixner, Christian Rainer and Kimmlingen, Ralph and Lanz, Titus and Schmitter, Sebastian and Uder, Michael and Nagel, Armin Michael},
title = {Accelerated B1+$$ \mathrmB_1^+ $$ mapping and robust parallel transmit pulse design for heart and prostate imaging at 7 T},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30185}
}
EUVIMED