EUVIMEDObjevte EUVIMED
Open access · CC BY 4.0

Modeling and measurement of lead tip heating and resonant length for implanted, insulated wires

Lydia J Bardwell Speltz iD, Seung-Kyun Lee, Yunhong Shu, Matthew R Tarasek, Joshua D Trzasko, Thomas K F Foo, Matt A Bernstein iD

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30145 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2024. Written by Lydia J Bardwell Speltz and 6 co-authors. Institution: Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA. Major MEDLINE topics: Hot Temperature. Cited 1 times so far (Crossref); 29 references. The full text is freely available (open access, CC BY 4.0). Funded by: NIBIB NIH HHS, NIH HHS, National Institutes of Health.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Purpose

To study implant lead tip heating because of the RF power deposition by developing mathematical models and comparing them with measurements acquired at 1.5 T and 3 T, especially to predict resonant length. Theory and Methods A simple exponential model and an adapted transmission line model for the electric field transfer function were developed. A set of wavenumbers, including that calculated from insulated antenna theory (King wavenumber) and that of the embedding medium were considered. Experiments on insulated, capped wires of varying lengths were performed to determine maximum temperature rise under RF exposure. The results are compared with model predictions from analytical expressions derived under the assumption of a constant electric field, and with those numerically calculated from spatially varying, simulated electric fields from body coil transmission. Simple expressions for the resonant length bounded between one‐quarter and one‐half wavelength are developed based on the roots of transcendental equations.

Results

The King wavenumber for both models more closely matched the experimental data with a maximum root mean square error of 9.81°C at 1.5 T and 5.71°C at 3 T compared to other wavenumbers with a maximum root mean square error of 27.52°C at 1.5 T and 22.01°C for 3 T. Resonant length was more accurately predicted compared to values solely based on the embedding medium.

Conclusion

Analytical expressions were developed for implanted lead heating and resonant lengths under specific assumptions. The value of the wavenumber has a strong effect on the model predictions. Our work could be used to better manage implanted device lead tip heating.

Subjects

MeSH terms (MEDLINE)

Hot TemperatureElectrodes, ImplantedEquipment DesignComputer SimulationModels, TheoreticalProstheses and ImplantsMagnetic Resonance ImagingEquipment Failure AnalysisHumans

Author keywords

MR safetyRF heatingimplanted devicesresonant lengthwavenumber

Authors and institutions

  • Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, Minnesota, USA.
  • Technology and Innovation Center, GE HealthCare, GE Research, Niskayuna, New York, USA.

ORCID profiles: Lydia J Bardwell Speltz, Matt A Bernstein

Funding

  • NIBIB NIH HHS
  • NIH HHS
  • National Institutes of Health

Cite

APA

Bardwell Speltz, L. J., Lee, S. K., Shu, Y., Tarasek, M. R., Trzasko, J. D., Foo, T. K. F., & Bernstein, M. A. (2024). Modeling and measurement of lead tip heating and resonant length for implanted, insulated wires. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30145

Vancouver

Bardwell Speltz LJ, Lee SK, Shu Y, Tarasek MR, Trzasko JD, Foo TKF, et al. Modeling and measurement of lead tip heating and resonant length for implanted, insulated wires. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30145

BibTeX
@article{BardwellSpeltz2024,
  author = {Bardwell Speltz, Lydia J and Lee, Seung-Kyun and Shu, Yunhong and Tarasek, Matthew R and Trzasko, Joshua D and Foo, Thomas K F and Bernstein, Matt A},
  title = {Modeling and measurement of lead tip heating and resonant length for implanted, insulated wires},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30145}
}
Start for free →