Design of multi-row parallel-transmit coil arrays for enhanced SAR efficiency with deep brain electrodes at 3T: an electromagnetic simulation study
Magnetic Resonance Materials in Physics, Biology and Medicine · 2024
DOI: 10.1007/s10334-024-01212-4 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2024. Verfasst von Nejat Karadeniz und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering and Imaging Science, King's College London, 3rd Floor Lambeth Wing, St Thomas' Hospital, Westminster Bridge Road, Lo…. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Deep Brain Stimulation, Brain, Electrodes, Implanted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: EPSRC Centre for Doctoral Training in Medical Imaging, Wellcome EPSRC Centre for Medical Engineering.
Abstract (Originalsprache)
Objective
Tissue heating near the implanted deep brain stimulation (DBS) during magnetic resonance imaging (MRI) poses a significant safety constraint. This study aimed to evaluate the performance of parallel transmit (pTx) head transmit radiofrequency (RF) coils in DBS patients, with a focus on excitation fidelity under specific absorption rate (SAR) control for brain imaging at 3T MRI. Materials and methods We employed electromagnetic simulations to assess different coil configurations, including multi-row pTx coils of 16–24 channels arranged in 1, 2, and 3 rows, and compared these to a circularly polarised and pTx birdcage coil using a realistic human model without and with DBS leads and electrodes.
Results
Two- and three-row pTx coils with overlapping loop elements exhibited similar performance, which was superior in excitation homogeneity and local SAR compared to the single-row coil and the birdcage coil both without and with DBS.
Discussion
These findings suggest that multi-row coils can enhance the safety and efficacy of MRI in patients with DBS devices, so potentially improving imaging performance in this expanding patient population. There was a minimal difference in performance between the 2 and 3-row coils, favouring the simpler, lower channel count design for practical implementation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Biomedical Engineering and Imaging Science, King's College London, 3rd Floor Lambeth Wing, St Thomas' Hospital, Westminster Bridge Road, London, SE1 7EH, UK. nejat.karadeniz@kcl.ac.uk.
- School of Biomedical Engineering and Imaging Science, King's College London, 3rd Floor Lambeth Wing, St Thomas' Hospital, Westminster Bridge Road, London, SE1 7EH, UK.
- Centre for the Developing Brain, King's College London, London, UK.
Förderung
- EPSRC Centre for Doctoral Training in Medical Imaging
- Wellcome EPSRC Centre for Medical Engineering
Zitieren
APA
Karadeniz, N., Hajnal, J. V., & Ipek, Ö. (2024). Design of multi-row parallel-transmit coil arrays for enhanced SAR efficiency with deep brain electrodes at 3T: an electromagnetic simulation study. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-024-01212-4
Vancouver
Karadeniz N, Hajnal JV, Ipek Ö. Design of multi-row parallel-transmit coil arrays for enhanced SAR efficiency with deep brain electrodes at 3T: an electromagnetic simulation study. Magnetic Resonance Materials in Physics, Biology and Medicine. 2024. doi:10.1007/s10334-024-01212-4
BibTeX
@article{Karadeniz2024,
author = {Karadeniz, Nejat and Hajnal, Joseph V and Ipek, Özlem},
title = {Design of multi-row parallel-transmit coil arrays for enhanced SAR efficiency with deep brain electrodes at 3T: an electromagnetic simulation study},
journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
year = {2024},
doi = {10.1007/s10334-024-01212-4}
}
EUVIMED