Characterization of concomitant gradient fields and their effects on image distortions using a low‐field point‐of‐care Halbach‐based MRI system

Bart de Vos iD, Rob F Remis, Andrew G Webb

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.29879 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Bart de Vos und 2 weiteren Autor:innen. Beteiligte Einrichtung: Radiology, C.J. Gorter MRI Center, Leiden University Medical Center, Leiden, The Netherlands. Hauptthemen laut MEDLINE: Point-of-Care Systems, Artifacts. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: H2020 European Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Concomitant gradient fields have been extensively studied at clinical field strengths. However, their effects have not yet been modeled for low‐field point‐of‐care (POC) systems. The purpose of this work is to characterize the effects associated with concomitant fields for POC Halbach‐array‐based systems.

Methods

The concomitant fields associated with a cylindrical gradient coils designed for a transverse and a signal model including the tilting effect of the effective magnetic field are derived. The formalism is used to simulate and predict concomitant field related distortions. A 46‐mT Halbach‐array‐based system with a maximum gradient strength of 15 mT/m is used to verify the model using two‐dimensional spin‐echo sequences.

Results

The simulations and experimental results are in good agreement with the derived equations. The fundamental characteristics of the concomitant field equations are different to conventional MRI systems: Image distortions occur primarily in the transverse directions and a cross‐term only exists when applying transverse gradient pulses simultaneously.

Conclusion

The level of image warping in the frequency encoding direction is insignificant for the POC systems discussed here. However, when trying to achieve short echo‐times by using strong phase encoding and readout‐dephasing gradients, the combination can result in image warping and blurring which should be accounted for in image interpretation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Point-of-Care SystemsArtifactsMagnetic Resonance ImagingMagnetic FieldsPhantoms, Imaging

Stichwörter der Autor:innen

Halbach-arrayconcomitant fieldsimage distortionslow-field

Autor:innen und Institutionen

  • Radiology, C.J. Gorter MRI Center, Leiden University Medical Center, Leiden, The Netherlands.
  • Micro-electronics, Signal Processing Systems, Delft University of Technology, Delft, The Netherlands.
  • Micro-electronics, Terahertz Sensing, Delft University of Technology, Delft, The Netherlands.

ORCID-Profile: Bart de Vos

Förderung

  • H2020 European Research Council

Zitieren

APA

de Vos, B., Remis, R. F., & Webb, A. G. (2024). Characterization of concomitant gradient fields and their effects on image distortions using a low‐field point‐of‐care Halbach‐based MRI system. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29879

Vancouver

de Vos B, Remis RF, Webb AG. Characterization of concomitant gradient fields and their effects on image distortions using a low‐field point‐of‐care Halbach‐based MRI system. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.29879

BibTeX
@article{deVos2024,
  author = {de Vos, Bart and Remis, Rob F and Webb, Andrew G},
  title = {Characterization of concomitant gradient fields and their effects on image distortions using a low‐field point‐of‐care Halbach‐based MRI system},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.29879}
}
Kostenlos starten →