Phase contrast MRI with minimized background phase errors

Michael Loecher iD, Daniel B Ennis

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30336 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Michael Loecher und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Stanford University, Stanford, California, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NHLBI NIH HHS, NIBIB NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Phase contrast MRI (PC‐MRI) is used clinically to measure velocities in the body, but systematic background phase errors caused by magnetic field imperfections corrupt the velocity measurements with offsets that limit clinical utility. This work aims to minimize systematic background phase errors in PC‐MRI, thereby maximizing the accuracy of velocity measurements.

Methods

The MRI scanner's background phase errors from eddy currents and mechanical oscillations were modeled using the gradient impulse response function (GIRF). Gradient waveforms were then numerically optimized using the GIRF to create pulse sequences that minimize the background phase errors. The pulse sequences were tested in a static phantom where the predicted response could be compared directly to the measured background velocity. The optimized acquisitions were then tested in human subjects, where flow rates and background errors were compared to conventional PC‐MRI.

Results

When using the GIRF‐optimized gradient waveforms, the predicted background phase was within 0.6 [95% CI = −3.4, 5.4] mm/s of the measured background phase in a static phantom. Excellent agreement was seen for in vivo blood flow values (flow rate agreement = 0.96), and the background phase was reduced by 78.8 18.7%.

Conclusion

This work shows that using a GIRF to model the effects of magnetic field imperfections combined with numerically optimized gradient waveforms enables PC‐MRI waveforms to be designed to produce a minimal background phase in the most time‐efficient manner. The methodology could be adapted for other MRI sequences where similar magnetic field errors affect measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMagnetic Resonance ImagingHumansBlood Flow VelocityAlgorithmsImage Processing, Computer-AssistedMagnetic FieldsReproducibility of ResultsArtifacts

Stichwörter der Autor:innen

background phasegradient impulse responsegradient optimizationphase contrast

Autor:innen und Institutionen

  • Department of Radiology, Stanford University, Stanford, California, USA.
  • Department of Radiology, VA Palo Alto Health Care System, Palo Alto, California, USA.

ORCID-Profile: Michael Loecher

Förderung

  • NHLBI NIH HHS
  • NIBIB NIH HHS
  • National Institutes of Health

Zitieren

APA

Loecher, M., & Ennis, D. B. (2025). Phase contrast MRI with minimized background phase errors. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30336

Vancouver

Loecher M, Ennis DB. Phase contrast MRI with minimized background phase errors. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30336

BibTeX
@article{Loecher2025,
  author = {Loecher, Michael and Ennis, Daniel B},
  title = {Phase contrast MRI with minimized background phase errors},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30336}
}
Kostenlos starten →