Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Radhika Tibrewala und 5 weiteren Autor:innen. Beteiligte Einrichtung: Center for Advanced Imaging Innovation and Research (CAI2R), Department of Radiology, New York University Grossman School of Medicine, New York, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, NINDS NIH HHS, NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Low‐cost and application‐specific scanners with unconventional designs can entail substantial inhomogeneities of the main magnetic field and non‐linearities of the gradient fields, which challenge assumptions made in MRI simulators. This work aims to develop a simulation framework that captures the encoding effects of strong field variations, accurately and efficiently, to enable the assessment of effects such as geometrical distortions, signal dropout, and foldover artifacts.

Methods

Like many other MRI simulators, ours discretizes magnetic fields in space. However, we extend the MR signal simulation at each grid point from the 0th‐order approximation, which assumes piecewise constant fields, to a 1st‐order approximation, which assumes piecewise linear fields. We solve the signal equation by analytically integrating over each grid cube, assuming linear field variations, and then summing over all cubes. We provide analytical integrals for several pulse sequences.

Results

The 1st‐order approximation captures strongly varying fields and associated intravoxel dephasing more accurately, avoiding severe “ringing” artifacts present in the usual 0th‐order simulations. This enables simulations on a much coarser grid, facilitating computational feasibility.

Conclusion

The first‐order simulator enables the evaluation of unconventional scanner designs with strongly varying magnetic fields.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage Processing, Computer-AssistedComputer SimulationAlgorithmsHumansArtifactsPhantoms, ImagingMagnetic FieldsReproducibility of Results

Stichwörter der Autor:innen

accessible MRIinhomogeneous fieldssimulationsspatial encoding

Autor:innen und Institutionen

  • Center for Advanced Imaging Innovation and Research (CAI2R), Department of Radiology, New York University Grossman School of Medicine, New York, USA.
  • Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University Grossman School of Medicine, New York, USA.
  • Vilcek Institute of Graduate Biomedical Sciences, Department of Radiology, New York University Grossman School of Medicine, New York, USA.
  • Siemens Healthcare GmbH, Erlangen, Bavaria, Germany.

ORCID-Profile: Radhika Tibrewala, Christopher M Collins, Daniel K Sodickson, Jakob Assländer

Förderung

  • NIBIB NIH HHS
  • NINDS NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Tibrewala, R., Collins, C. M., Mallett, M., Vom Endt, A., Sodickson, D. K., & Assländer, J. (2026). First‐Order Spatial Encoding Simulations for Improved Accuracy in the Presence of Strong B 0 and Gradient Field Variations. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70160

Vancouver

Tibrewala R, Collins CM, Mallett M, Vom Endt A, Sodickson DK, Assländer J. First‐Order Spatial Encoding Simulations for Improved Accuracy in the Presence of Strong B 0 and Gradient Field Variations. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70160

BibTeX
@article{Tibrewala2026,
  author = {Tibrewala, Radhika and Collins, Christopher M and Mallett, Michael and Vom Endt, Axel and Sodickson, Daniel K and Assländer, Jakob},
  title = {First‐Order Spatial Encoding Simulations for Improved Accuracy in the Presence of Strong B 0 and Gradient Field Variations},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  doi = {10.1002/mrm.70160}
}
Kostenlos starten →