A correction algorithm for improved magnetic field monitoring with distal field probes

Paul I Dubovan iD, Kyle M Gilbert iD, Corey A Baron iD

Magnetic Resonance in Medicine · 2023

DOI: 10.1002/mrm.29781 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2023. Verfasst von Paul I Dubovan und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Biophysics, Western University, London, Ontario, Canada. Hauptthemen laut MEDLINE: Algorithms, Artifacts, Image Processing, Computer-Assisted, Magnetic Resonance Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Canada First Research Excellence Fund to BrainsCAN, Canada Foundation for Innovation, Canada Research Chairs.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

A significant source of artifacts in MRI are field fluctuations. Field monitoring is a new technology that allows measurement of field dynamics during a scan via “field probes,” which can be used to improve image reconstruction. Ideally, probes are located within the volume where gradients produce nominally linear field patterns. However, in some situations probes must be located far from isocenter where rapid field variation can arise, leading to erroneous field‐monitoring characterizations and images. This work aimed to develop an algorithm that improves the robustness of field dynamics in these situations.

Methods

The algorithm is split into three components. Component 1 calculates field dynamics one spatial order at a time, whereas the second implements a weighted least squares solution based on probe distance. Component 3 then calculates phase residuals and removes the residual phase for distant probes before recalculation. Two volunteers and a phantom were scanned on a 7T MRI using diffusion‐weighted sequences, and field monitoring was performed. Image reconstructions were informed with field dynamics calculated conventionally, and with the correction algorithm, after which in vivo images were compared qualitatively and phantom image error was quantitatively assessed.

Results

The algorithm was able to correct corrupted field dynamics, resulting in image‐quality improvements. Significant artifact reduction was observed when correcting higher‐order fits. Stepwise fitting provided the most correction benefit, which was marginally improved when adding the other correction strategies.

Conclusion

The proposed algorithm can mitigate effects of phase errors in field monitoring, providing improved characterization of field dynamics.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsArtifactsImage Processing, Computer-AssistedMagnetic Resonance ImagingBrainHumansPhantoms, ImagingMagnetic Fields

Stichwörter der Autor:innen

eddy currentsfield monitoringfield probesgradient nonlinearityhigher order reconstructionradiofrequency coil

Autor:innen und Institutionen

  • Department of Medical Biophysics, Western University, London, Ontario, Canada.
  • Center for Functional and Metabolic Mapping, Western University, London, Ontario, Canada.

ORCID-Profile: Paul I Dubovan, Kyle M Gilbert, Corey A Baron

Förderung

  • Canada First Research Excellence Fund to BrainsCAN
  • Canada Foundation for Innovation
  • Canada Research Chairs
  • NSERC Discovery Grant
  • NSERC PGS D Program
  • Ontario Research Fund

Zitieren

APA

Dubovan, P. I., Gilbert, K. M., & Baron, C. A. (2023). A correction algorithm for improved magnetic field monitoring with distal field probes. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29781

Vancouver

Dubovan PI, Gilbert KM, Baron CA. A correction algorithm for improved magnetic field monitoring with distal field probes. Magnetic Resonance in Medicine. 2023. doi:10.1002/mrm.29781

BibTeX
@article{Dubovan2023,
  author = {Dubovan, Paul I and Gilbert, Kyle M and Baron, Corey A},
  title = {A correction algorithm for improved magnetic field monitoring with distal field probes},
  journal = {Magnetic Resonance in Medicine},
  year = {2023},
  doi = {10.1002/mrm.29781}
}
Kostenlos starten →