Computational methods for the estimation of ideal current patterns in realistic human models
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.29864 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Ilias I Giannakopoulos und 3 weiteren Autor:innen. Beteiligte Einrichtung: The Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University Grossman School of Medicine, New York, New…. Hauptthemen laut MEDLINE: Electromagnetic Fields, Magnetic Resonance Imaging. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 64 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
To introduce a method for the estimation of the ideal current patterns (ICP) that yield optimal signal‐to‐noise ratio (SNR) for realistic heterogeneous tissue models in MRI. Theory and Methods The ICP were calculated for different surfaces that resembled typical radiofrequency (RF) coil formers. We constructed numerical electromagnetic (EM) bases to accurately represent EM fields generated by RF current sources located on the current‐bearing surfaces. Using these fields as excitations, we solved the volume integral equation and computed the EM fields in the sample. The fields were appropriately weighted to calculate the optimal SNR and the corresponding ICP. We demonstrated how to qualitatively use ICP to guide the design of a coil array to maximize SNR inside a head model.
Results
In agreement with previous analytic work, ICP formed large distributed loops for voxels in the middle of the sample and alternated between a single loop and a figure‐eight shape for a voxel 3‐cm deep in the sample's cortex. For the latter voxel, a surface quadrature loop array inspired by the shape of the ICP reached of the optimal SNR at 3T, whereas a single loop placed above the voxel reached only of the optimal SNR. At 7T, the performance of the two designs decreased to and , respectively, suggesting that loops could be suboptimal at ultra‐high field MRI.
Conclusion
ICP can be calculated for human tissue models, potentially guiding the design of application‐specific RF coil arrays.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The Bernard and Irene Schwartz Center for Biomedical Imaging, Department of Radiology, New York University Grossman School of Medicine, New York, New York, USA.
- R&D Department, Corsmed, Stockholm, Sweden.
- Center for Advanced Imaging Innovation and Research (CAI2R), Department of Radiology, New York University Grossman School of Medicine, New York, New York, USA.
ORCID-Profile: Ilias I Giannakopoulos, Ioannis P Georgakis, Daniel K Sodickson, Riccardo Lattanzi
Förderung
- NIBIB NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Giannakopoulos, I. I., Georgakis, I. P., Sodickson, D. K., & Lattanzi, R. (2024). Computational methods for the estimation of ideal current patterns in realistic human models. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.29864
Vancouver
Giannakopoulos II, Georgakis IP, Sodickson DK, Lattanzi R. Computational methods for the estimation of ideal current patterns in realistic human models. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.29864
BibTeX
@article{Giannakopoulos2024,
author = {Giannakopoulos, Ilias I and Georgakis, Ioannis P and Sodickson, Daniel K and Lattanzi, Riccardo},
title = {Computational methods for the estimation of ideal current patterns in realistic human models},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.29864}
}
EUVIMED