EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Signal-to-noise trade-offs between magnet diameter and shield-to-coil distance for cylindrical Halbach-based portable MRI systems for neuroimaging

Javad Parsa, Andrew Webb

Magnetic Resonance Materials in Physics, Biology and Medicine · 2024

DOI: 10.1007/s10334-024-01210-6 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance Materials in Physics, Biology and Medicine, 2024. Verfasst von Javad Parsa und 1 weiteren Autor:innen. Beteiligte Einrichtung: C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands. javadparsa1m@gmail.com. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Neuroimaging, Magnets. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: H2020 European Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To investigate the trade-off between magnet bore diameter and the distance between the conductive Faraday shield and RF head coil for low-field point-of-care neuroimaging systems.

Methods

Electromagnetic simulations were performed for three different Faraday shield geometries and two commonly used RF coil designs (spiral and solenoid) to assess the effects of a close-fitting shield on the RF coil's transmit and receive efficiencies. Experimental measurements were performed to confirm the accuracy of the simulations. Parallel simulations were performed to assess the static magnet ( $${B}_{0}$$ B 0 ) field as a function of the magnet bore diameter. The obtainable SNR was then calculated as a function of these two related variables.

Results

Simulations of the RF coil characteristics and $${B}_{1}^{+}$$ B 1 + transmit efficiencies agreed well with corresponding experimentally determined parameters. Overall, the RF coil transmit efficiency was, as expected, higher when the gap between the shield and coil increased. The calculated intrinsic SNR showed that maximum SNR would be obtained for a cylindrical shield of diameter 310 mm with an inner diameter of the magnet of 320 mm (assuming 10 mm for the gradient coils).

Conclusion

This work presents an overview of the trade-offs in transmit efficiencies for RF coils used for POC MRI neuroimaging as a function of coil-to-shield distance and inner diameter of the Halbach magnet. Results show that there is a relatively shallow optimum between a magnet diameter of 290 and 330 mm, with values falling more than 10% if either smaller or larger magnets are used.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingNeuroimagingMagnetsSignal-To-Noise RatioHumansEquipment DesignComputer SimulationPhantoms, ImagingRadio WavesPoint-of-Care SystemsElectromagnetic Fields

Stichwörter der Autor:innen

Helbach-array MRI magnetIntrinsic SNRRF coilSimulationUltra Low field MRI

Autor:innen und Institutionen

  • C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands. javadparsa1m@gmail.com.
  • C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center, Leiden, The Netherlands.

Förderung

  • H2020 European Research Council

Zitieren

APA

Parsa, J., & Webb, A. (2024). Signal-to-noise trade-offs between magnet diameter and shield-to-coil distance for cylindrical Halbach-based portable MRI systems for neuroimaging. Magnetic Resonance Materials in Physics, Biology and Medicine. https://doi.org/10.1007/s10334-024-01210-6

Vancouver

Parsa J, Webb A. Signal-to-noise trade-offs between magnet diameter and shield-to-coil distance for cylindrical Halbach-based portable MRI systems for neuroimaging. Magnetic Resonance Materials in Physics, Biology and Medicine. 2024. doi:10.1007/s10334-024-01210-6

BibTeX
@article{Parsa2024,
  author = {Parsa, Javad and Webb, Andrew},
  title = {Signal-to-noise trade-offs between magnet diameter and shield-to-coil distance for cylindrical Halbach-based portable MRI systems for neuroimaging},
  journal = {Magnetic Resonance Materials in Physics, Biology and Medicine},
  year = {2024},
  doi = {10.1007/s10334-024-01210-6}
}
Kostenlos starten →