Diffusion posterior sampling framework for joint image reconstruction and scatter estimation in cone‐beam CT

Altea Lorenzon, Xiao Jiang, Grace J Gang, J Webster Stayman

Medical Physics · 2026-11 · 53 (11)

DOI: 10.1002/mp.70755 ↗

At a glance

Journal article in Medical Physics, published online on 2026-11-11. Written by Altea Lorenzon and 3 co-authors. Institution: Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland, USA. Major MEDLINE topics: Cone-Beam Computed Tomography, Scattering, Radiation, Image Processing, Computer-Assisted. 76 references. Funded by: NCI NIH HHS, NIH, National Institutes of Health.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

X‐ray scatter can have a significant impact on image quality in cone beam CT (CBCT) including the introduction of severe artifacts and CT value inaccuracies. Classic solutions to the scatter problem often involve specialized hardware and data acquisition, or computationally expensive estimation of scatter via Monte Carlo (MC) methods.

Purpose

We explore a novel scatter correction approach that leverages the recent development of Diffusion Posterior Sampling (DPS) to jointly estimate CT attenuation and scattered radiation directly from single‐energy CT projection data.

Methods

We rewrite the nonlinear forward model for CT measurements to include an unknown additive scatter term which we jointly estimate with the image volume. Solution of this ill‐posed problem is enabled by diffusion models that capture the distribution of scatter‐free images of anatomy. The joint reconstruction and scatter estimation is evaluated using both simulation and phantom data and compared with a MC‐based scatter correction method.

Results

Qualitative and quantitative results obtained from simulated and phantom data shows that the proposed approach can effectively estimate the scatter contribution, significantly remove scatter artifacts, and restore the overall image contrast and attenuation values.

Conclusions

Joint estimation of scatter and reconstruction for CBCT has the potential to provide improved CBCT image quality and quantitation through direct estimation without the addition of specialized hardware or MC simulations.

Subjects

MeSH terms (MEDLINE)

Cone-Beam Computed TomographyScattering, RadiationImage Processing, Computer-AssistedPhantoms, ImagingDiffusionMonte Carlo MethodArtifacts

Author keywords

cone‐beam computed tomography (CBCT)diffusion posterior samplingscatter estimation

Authors and institutions

  • Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland, USA.
  • Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.

Funding

  • NCI NIH HHS
  • NIH
  • National Institutes of Health

Cite

APA

Lorenzon, A., Jiang, X., Gang, G. J., & Stayman, J. W. (2026). Diffusion posterior sampling framework for joint image reconstruction and scatter estimation in cone‐beam CT. Medical Physics, 53(11). https://doi.org/10.1002/mp.70755

Vancouver

Lorenzon A, Jiang X, Gang GJ, Stayman JW. Diffusion posterior sampling framework for joint image reconstruction and scatter estimation in cone‐beam CT. Medical Physics. 2026;53(11). doi:10.1002/mp.70755

BibTeX
@article{Lorenzon2026,
  author = {Lorenzon, Altea and Jiang, Xiao and Gang, Grace J and Stayman, J Webster},
  title = {Diffusion posterior sampling framework for joint image reconstruction and scatter estimation in cone‐beam CT},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {11},
  doi = {10.1002/mp.70755}
}
Start for free →