A rotating beam‐blocker method for cone beam CT scatter correction
Medical Physics · 2024-10 · 51 (10) · 7320-7331
DOI: 10.1002/mp.17274 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 10.07.2024. Verfasst von Hehe Cui und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Engineering and Applied Physics, University of Science and Technology of China, Hefei, Anhui, People's Republic of China. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Scattering, Radiation, Artifacts. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 43 Literaturangaben. Gefördert von: Ministry of Science and Technology of China Key Research and Development Project, Fundamental Research Funds for the Central Universities, Hefei Comprehensive National Science Center.
Abstract (Originalsprache)
Background
Cone beam CT (CBCT) is widely utilized in clinics. However, the scatter artifact degrades the CBCT image quality, hampering the expansion of CBCT applications. Recently, beam‐blocker methods have been used for CBCT scatter correction and proved their high cost‐effectiveness.
Purpose
A rotating beam‐blocker (RBB) method for CBCT scatter correction was proposed to complete scatter correction and image reconstruction within a single scan in both full‐ and half‐fan scan scenarios.
Methods
The RBB consisted of two open regions and two blocked regions, and was designed as a centrosymmetric structure. The open and blocked projections could be alternatively obtained within one single rotation. The open projections were corrected with the scatter signal calculated from the blocked projections, and then used to reconstruct the 3D image via the Feldkamp‐Davis‐Kress algorithm. The performance of the RBB method was evaluated on head and pelvis phantoms in scenarios with and without a bowtie filter. The images obtained from nine repeated scans in each scenario were used to calculate the evaluation metrics including the CT number error, spatial nonuniformity (SNU) and contrast‐to‐noise ratio (CNR).
Results
For the head phantom, the CT number error was decreased to 200 HU before correction when scanned without a bowtie filter, and to 160 HU when scanned with a full bowtie filter. For the pelvis phantom, the CT number error was reduced to 250 HU before correction when scanned without a bowtie filter, and to 190 HU when scanned with a half bowtie filter. After scatter correction, the uniformity and contrast were both improved, resulting in an SNU of >79% decrease and CNR of >2 times increase, respectively.
Conclusions
High‐quality CBCT images could be obtained in a single scan after using the proposed RBB method for scatter correction, enabling more accurate image guidance for surgery and radiation therapy applications. With almost no time delay between the successive open and blocked projections, the RBB method could eliminate the motion‐induced anatomical mismatches between the corresponding open and blocked projections and could find particular usefulness in thoracic and abdominal imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Engineering and Applied Physics, University of Science and Technology of China, Hefei, Anhui, People's Republic of China.
- Cancer Center, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, People's Republic of China.
- Institute of Radiation Oncology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, People's Republic of China.
- Hubei Key Laboratory of Precision Radiation Oncology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, People's Republic of China.
- Department of Radiation Oncology, the First Affiliated Hospital of USTC, Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei, Anhui, People's Republic of China.
- Ion Medical Research Institute, University of Science and Technology of China, Hefei, Anhui, People's Republic of China.
Förderung
- Ministry of Science and Technology of China Key Research and Development Project
- Fundamental Research Funds for the Central Universities
- Hefei Comprehensive National Science Center
Zitieren
APA
Cui, H., Zhan, H., Yuan, X., & Yang, Y. (2024). A rotating beam‐blocker method for cone beam CT scatter correction. Medical Physics, 51(10), 7320-7331. https://doi.org/10.1002/mp.17274
Vancouver
Cui H, Zhan H, Yuan X, Yang Y. A rotating beam‐blocker method for cone beam CT scatter correction. Medical Physics. 2024;51(10):7320-7331. doi:10.1002/mp.17274
BibTeX
@article{Cui2024,
author = {Cui, Hehe and Zhan, Haolin and Yuan, Xiaogang and Yang, Yidong},
title = {A rotating beam‐blocker method for cone beam CT scatter correction},
journal = {Medical Physics},
year = {2024},
volume = {51},
number = {10},
pages = {7320--7331},
doi = {10.1002/mp.17274}
}
EUVIMED