Diffusion posterior sampling framework for joint image reconstruction and scatter estimation in cone‐beam CT
Medical Physics · 2026-11 · 53 (11)
DOI: 10.1002/mp.70755 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 11.11.2026. Verfasst von Altea Lorenzon und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland, USA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Scattering, Radiation, Image Processing, Computer-Assisted. Die Arbeit nennt 76 Literaturangaben. Gefördert von: NCI NIH HHS, NIH, National Institutes of Health.
Abstract (Originalsprache)
Background
X‐ray scatter can have a significant impact on image quality in cone beam CT (CBCT) including the introduction of severe artifacts and CT value inaccuracies. Classic solutions to the scatter problem often involve specialized hardware and data acquisition, or computationally expensive estimation of scatter via Monte Carlo (MC) methods.
Purpose
We explore a novel scatter correction approach that leverages the recent development of Diffusion Posterior Sampling (DPS) to jointly estimate CT attenuation and scattered radiation directly from single‐energy CT projection data.
Methods
We rewrite the nonlinear forward model for CT measurements to include an unknown additive scatter term which we jointly estimate with the image volume. Solution of this ill‐posed problem is enabled by diffusion models that capture the distribution of scatter‐free images of anatomy. The joint reconstruction and scatter estimation is evaluated using both simulation and phantom data and compared with a MC‐based scatter correction method.
Results
Qualitative and quantitative results obtained from simulated and phantom data shows that the proposed approach can effectively estimate the scatter contribution, significantly remove scatter artifacts, and restore the overall image contrast and attenuation values.
Conclusions
Joint estimation of scatter and reconstruction for CBCT has the potential to provide improved CBCT image quality and quantitation through direct estimation without the addition of specialized hardware or MC simulations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland, USA.
- Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
Förderung
- NCI NIH HHS
- NIH
- National Institutes of Health
Zitieren
APA
Lorenzon, A., Jiang, X., Gang, G. J., & Stayman, J. W. (2026). Diffusion posterior sampling framework for joint image reconstruction and scatter estimation in cone‐beam CT. Medical Physics, 53(11). https://doi.org/10.1002/mp.70755
Vancouver
Lorenzon A, Jiang X, Gang GJ, Stayman JW. Diffusion posterior sampling framework for joint image reconstruction and scatter estimation in cone‐beam CT. Medical Physics. 2026;53(11). doi:10.1002/mp.70755
BibTeX
@article{Lorenzon2026,
author = {Lorenzon, Altea and Jiang, Xiao and Gang, Grace J and Stayman, J Webster},
title = {Diffusion posterior sampling framework for joint image reconstruction and scatter estimation in cone‐beam CT},
journal = {Medical Physics},
year = {2026},
volume = {53},
number = {11},
doi = {10.1002/mp.70755}
}
EUVIMED