Development of an adjustable dynamic phantom for testing deviceless motion correction algorithm in PET
Medical Physics · 2026-08 · 53 (8)
DOI: 10.1002/mp.70577 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 21.07.2026. Verfasst von Christian Kühnel und 6 weiteren Autor:innen. Beteiligte Einrichtung: Clinic for Nuclear Medicine, University Hospital Jena, Jena, Thuringia, Germany. Hauptthemen laut MEDLINE: Phantoms, Imaging, Algorithms, Image Processing, Computer-Assisted, Positron-Emission Tomography. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Respiratory motion during PET Imaging leads to blurring, volumetric overestimation, and reduced quantitative accuracy. Deviceless motion correction algorithms (MCA) such as OncoFreeze (OF) by Siemens Healthineers (Forchheim, Germany) extract respiratory waveforms from list‐mode data but require validation that cannot be obtained in vivo.
Purpose
To develop and validate an adjustable dynamic phantom with fillable spheres for quantitative assessment of deviceless PET motion correction.
Methods
A microprocessor‐controlled two‐axis phantom was constructed to generate linear craniocaudal (cc), ventrodorsal (vd), and diagonal (diag) motion with amplitudes of 0–40 mm and frequencies of 10–40 min − 1 . Polymethylmethacrylate (PMMA) spheres (4, 1, and 0.25 mL) filled with 1 8 F‐fluorodeoxyglucose (FDG) were placed in a water‐filled PMMA tank and imaged on a Biograph Vision 600 PET/CT (Siemens Healthineers AG, Forchheim, Germany). For each sphere and direction, 2‐min PET scans were acquired with and without MCA, yielding 288 datasets. Volumes, activities, and recovery coefficients (RCs) were extracted using two different isocontours.
Results
Without correction, motion produced substantial overestimation of apparent volume: +15%–55% (4 mL), +81%–166% (1 mL) and +276%–509% (0.25 mL), depending on trajectory and IC 37 . MCA reduced these distortions: For the 1 mL sphere from +87% to −8% (vd) and +166% to +82% (diag). IC 0.1 ‐based activity remained within ≈1%–2% of the nominal value and was minimally affected by MCA, while IC 37 underestimated activity (∼45%) and was direction‐dependently altered by correction. RCs increased with sphere size and were partially improved by MCA but not restored to static levels.
Conclusions
The phantom provides a reproducible ground‐truth framework for evaluating deviceless PET motion correction. MCA reduces motion‐induced blurring and volumetric inflation, though correction remains size‐ and direction‐dependent. The system enables systematic benchmarking and further optimization of motion‐correction algorithms.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Clinic for Nuclear Medicine, University Hospital Jena, Jena, Thuringia, Germany.
- Central Research Facilities, University Hospital Jena, Jena, Thuringia, Deutschland.
Zitieren
APA
Kühnel, C., Schmidt, T. N., Schreiber, L., Ginter, H., Händel, M., Freesmeyer, M., & Gühne, F. (2026). Development of an adjustable dynamic phantom for testing deviceless motion correction algorithm in PET. Medical Physics, 53(8). https://doi.org/10.1002/mp.70577
Vancouver
Kühnel C, Schmidt TN, Schreiber L, Ginter H, Händel M, Freesmeyer M, et al. Development of an adjustable dynamic phantom for testing deviceless motion correction algorithm in PET. Medical Physics. 2026;53(8). doi:10.1002/mp.70577
BibTeX
@article{Khnel2026,
author = {Kühnel, Christian and Schmidt, Tabea Nikola and Schreiber, Leonie and Ginter, Holger and Händel, Michael and Freesmeyer, Martin and Gühne, Falk},
title = {Development of an adjustable dynamic phantom for testing deviceless motion correction algorithm in PET},
journal = {Medical Physics},
year = {2026},
volume = {53},
number = {8},
doi = {10.1002/mp.70577}
}
EUVIMED