A synergistic framework for geometric calibration and reconstruction in dual‐arm robotic cone‐beam computed tomography

Zeyu Lai, Shousheng Luo, Xiaohong Fan, Shijie Wang, Xiaoqin Xia, Dexing Kong

Medical Physics · 2026-07 · 53 (7)

DOI: 10.1002/mp.70558 ↗

At a glance

Journal article in Medical Physics, published online on 2026-07-10. Written by Zeyu Lai and 5 co-authors. Institution: School of Mathematical Sciences, Zhejiang University, Hangzhou, China. Major MEDLINE topics: Cone-Beam Computed Tomography, Robotics, Image Processing, Computer-Assisted. 51 references. The full text is freely available in PubMed Central.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

Dual‐arm robotic cone‐beam computed tomography (CBCT) systems are susceptible to geometric instability due to their decoupled kinematics, which can limit their practical application.

Purpose

This study aims to develop and evaluate a synergistic framework for improving geometric calibration and reconstruction consistency in flexible dual‐arm robotic CBCT platforms.

Methods

We introduce the Synergistic Calibration and Reconstruction (SyCaR) framework. It uses a two‐stage geometric calibration strategy that combines real‐time external metrology for projection‐wise pose estimation with an optional reference‐free projection‐consistency refinement for residual geometric correction. For reconstruction, we developed a Pose‐Driven Feldkamp‐Davis‐Kress (PD‐FDK) algorithm that operates directly on per‐projection pose data to account for detector misalignments and non‐ideal source‐detector trajectory deviations.

Results

Physical phantom experiments showed that encoder geometry produced severe streaking artifacts and structural inconsistency, whereas motion‐capture geometry restored clearer phantom structures. In the numerical study, PD‐FDK achieved the best PSNR and RMSE among the evaluated analytical methods, improving from 28.99 dB and 0.036 with standard FDK to 32.72 dB and 0.023. Across numerical and physical experiments, PD‐FDK reduced geometry‐related artifacts and showed consistent performance among FDK type methods under the tested conditions.

Conclusions

The proposed SyCaR framework provides a feasible computational approach for mitigating geometric instability in dual‐arm robotic CBCT. The combination of geometric calibration and PD‐FDK reconstruction improved reconstruction consistency and reduced geometry‐related artifacts under the evaluated numerical and physical settings, while retaining the computational efficiency of analytical reconstruction. This work supports the further development of flexible dual‐arm robotic CBCT for biomedical imaging applications, although broader validation with more diverse trajectories, objects, and acquisition conditions remains necessary.

Subjects

MeSH terms (MEDLINE)

Cone-Beam Computed TomographyRoboticsImage Processing, Computer-AssistedCalibrationPhantoms, ImagingAlgorithmsArtifacts

Author keywords

cone beam CTgeometric calibrationrobotic CT

Authors and institutions

  • School of Mathematical Sciences, Zhejiang University, Hangzhou, China.
  • College of Mathematical Medicine, Zhejiang Normal University, Jinhua, China.

Cite

APA

Lai, Z., Luo, S., Fan, X., Wang, S., Xia, X., & Kong, D. (2026). A synergistic framework for geometric calibration and reconstruction in dual‐arm robotic cone‐beam computed tomography. Medical Physics, 53(7). https://doi.org/10.1002/mp.70558

Vancouver

Lai Z, Luo S, Fan X, Wang S, Xia X, Kong D. A synergistic framework for geometric calibration and reconstruction in dual‐arm robotic cone‐beam computed tomography. Medical Physics. 2026;53(7). doi:10.1002/mp.70558

BibTeX
@article{Lai2026,
  author = {Lai, Zeyu and Luo, Shousheng and Fan, Xiaohong and Wang, Shijie and Xia, Xiaoqin and Kong, Dexing},
  title = {A synergistic framework for geometric calibration and reconstruction in dual‐arm robotic cone‐beam computed tomography},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {7},
  doi = {10.1002/mp.70558}
}
Start for free →