Real‐time energy measurement of clinical carbon ion beams using a cross‐correlation time‐of‐flight method with parallel‐plate chambers
Medical Physics · 2026-04 · 53 (4)
DOI: 10.1002/mp.70391 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 29.03.2026. Verfasst von Na Hye Kwon und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, Republic …. Hauptthemen laut MEDLINE: Heavy Ion Radiotherapy, Radiometry, Carbon. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Korea Institute for Advancement of Technology, National Research Foundation of Korea, Korea Foundation Of Nuclear Safety (KOFONS).
Abstract (Originalsprache)
Background
In carbon‐ion radiotherapy (CIRT), the beam energy determines both the particle range and the overall dosimetric quality. Range‐verification QA devices such as Zebra and Giraffe, which are based on multilayer ionization chambers (MLICs), can verify the range but only under dedicated QA conditions, leaving any energy deviations introduced by nozzle components undetected in real time. In particular, nozzle structures such as ridge filters can broaden or modulate the energy spectrum, causing the effective energy delivered to the patient to differ from the nominal accelerator setting. These limitations highlight the need for a real‐time method capable of verifying the beam energy under actual clinical operating conditions.
Purpose
We proposed a TOF‐based beam‐energy measurement concept that leverages a cross‐correlation analysis of full detector waveforms. Compact and radiation‐hard parallel‐plate chambers (PPCs) were developed and evaluated, in contrast to prior TOF systems based on semiconductor detectors.
Methods
PPCs (2.5 cm diameter active area, 0.4 mm gas gap) were operated in CO 2 . Two detectors were mounted coaxially with detector separations of 22.5 and 46.3 cm. Experiments were performed at Yonsei Heavy‐ion Therapy Center (HITC) using four nominal energies (102.6, 140.4, 250.3, 430 MeV/nucleon) and three intensities, covering the clinically interesting ranges. Signals were digitized with a 1 GHz bandwidth oscilloscope. For each spill, paired waveforms were cross‐correlated, and peak times were refined by parabolic interpolation to determine TOF. Precision and accuracy were evaluated across energies, intensities, and detector separations.
Results
The PPCs operated stably for all beam conditions. Under pencil‐beam delivery and normalized to 1 s acquisitions, the timing precision of the mean TOF (standard error) remained within 1 ps for both detector separations, scaling with (N: number of TOF samples per acquisition) and not representing the single‐particle TOF resolution. Residuals between measured and theoretical TOF remained within 80 ps across energies and distances. After relativistic conversion from TOF to kinetic energy and then to water‐equivalent range, all deviations were within a 1 mm range shift, meeting the recommended clinical criteria for range verification.
Conclusions
We demonstrated that compact CO 2 ‐filled PPCs, operated as a TOF pair, can measure carbon‐ion beam energy across the clinically relevant range of energies (≈100–430 MeV/u) and intensities used in routine treatment delivery. We achieved sub‐picosecond timing precision on the TOF mean (standard error) per 1 s acquisition and submillimeter water‐equivalent range accuracy using a robust cross‐correlation analysis method. These results open the way to the integration of PPC‐based TOF monitoring to tighten beam‐delivery tolerances and improve the reliability and safety of carbon‐ion radiotherapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine, Seoul, Republic of Korea.
- Oncosoft Inc., Seoul, South Korea.
- LIP - Laboratory of Instrumentation and Experimental Particle Physics, Coimbra, Portugal.
- ARA Inc., Seoul, South Korea.
- Coimbra Institute of Engineering, Polytechnic University of Coimbra, Coimbra, Portugal.
Förderung
- Korea Institute for Advancement of Technology
- National Research Foundation of Korea
- Korea Foundation Of Nuclear Safety (KOFONS)
Zitieren
APA
Kwon, N. H., Choi, S. W., Han, S., Yun, Y., Han, M. C., Hong, C. S., Kim, H. J., Lee, H., Kim, C., Kim, D. W., Koom, W. S., Kim, J. S., Carolino, N., Lopes, L., Kim, D. W., & Fonte, P. J. R. (2026). Real‐time energy measurement of clinical carbon ion beams using a cross‐correlation time‐of‐flight method with parallel‐plate chambers. Medical Physics, 53(4). https://doi.org/10.1002/mp.70391
Vancouver
Kwon NH, Choi SW, Han S, Yun Y, Han MC, Hong CS, et al. Real‐time energy measurement of clinical carbon ion beams using a cross‐correlation time‐of‐flight method with parallel‐plate chambers. Medical Physics. 2026;53(4). doi:10.1002/mp.70391
BibTeX
@article{Kwon2026,
author = {Kwon, Na Hye and Choi, Sung Woon and Han, Soorim and Yun, Yongdo and Han, Min Cheol and Hong, Chae-Seon and Kim, Ho Jin and Lee, Ho and Kim, Changhwan and Kim, Do Won and Koom, Woong Sub and Kim, Jin Sung and Carolino, Nuno and Lopes, Luis and Kim, Dong Wook and Fonte, Paulo J R},
title = {Real‐time energy measurement of clinical carbon ion beams using a cross‐correlation time‐of‐flight method with parallel‐plate chambers},
journal = {Medical Physics},
year = {2026},
volume = {53},
number = {4},
doi = {10.1002/mp.70391}
}
EUVIMED