Monte Carlo–calculated beam–quality and perturbation correction factors for helium–, carbon–, and neon–ion beams
Medical Physics · 2025
DOI: 10.1002/mp.70024 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Yuka Urago und 4 weiteren Autor:innen. Beteiligte Einrichtung: QST Hospital, National Institutes for Quantum Science and Technology (QST), Chiba, Japan. Hauptthemen laut MEDLINE: Monte Carlo Method, Helium, Neon, Carbon. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 39 Literaturangaben.
Abstract (Originalsprache)
Purpose
A new technique, multi‐ion radiotherapy, which uses He, O, and Ne ions in addition to C ions, has been proposed, though there is still limited information on the dosimetric quantities, such as the beam quality and perturbation correction factors, necessary to determine the absorbed dose‐to‐water for various ion species. In this study, beam‐quality correction factors, k Q , for He, C, and Ne ions in five types of ionization chambers were derived through Monte Carlo simulations. Additionally, perturbation effects attributed to the ionization chambers were analyzed in detail, and differences in these effects across chamber types and ions were clarified.
Methods
The k Q values for 150 MeV/u‐He, 290 MeV/u‐C, and 400 MeV/u‐Ne ions were evaluated using the Geant4 Monte Carlo code. The f Q (the product of the water‐to‐air stopping power ratio and the perturbation correction factor) was obtained using the absorbed doses to water and to air inside the active volume of the ionization chamber. The k Q values were then calculated from the f Q and the literature‐extracted factors, and W air . To evaluate individual perturbation effects, each chamber component (such as the central electrode, wall, and stem in the modeled chambers) was replaced by water in turn, and the dose was recalculated for each modified geometry. From the ratio of these doses, the contribution of each component to the perturbation effect was estimated.
Results
The derived k Q values were consistent with the code of practice for dosimetry, the revised TRS‐398, and varied slightly with the ion species, from He to Ne ions. Additionally, the k Q values increased by up to 1% compared with the current protocol values, particularly in plane‐parallel chambers. Analysis of contributions to the perturbation factors revealed that the perturbation effect due to the cavity, P cav , is significantly greater for light ions than for protons and had the most substantial impact among the evaluated individual perturbation factors.
Conclusions
The revision of k Q values for light ions may result in the absorbed dose‐to‐water varying by up to 1% compared with doses calculated using the currently recommended k Q values. At the current stage, it has been confirmed that common k Q values can be applied to light ions from He to Ne in both Farmer and Markus type ionization chambers.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- QST Hospital, National Institutes for Quantum Science and Technology (QST), Chiba, Japan.
- National Institute of Advanced Industrial Science and Technology (AIST), National Metrology Institute of Japan, Ibaraki, Japan.
- Department of Radiological Science, Graduate School of Human Health Science, Tokyo Metropolitan University, Tokyo, Japan.
- Division of Health Sciences, Osaka University, Osaka, Japan.
- School of Physics, University of Bristol, Bristol, UK.
- Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
Zitieren
APA
Urago, Y., Sakama, M., Sakata, D., Katayose, T., & Chang, W. (2025). Monte Carlo–calculated beam–quality and perturbation correction factors for helium–, carbon–, and neon–ion beams. Medical Physics. https://doi.org/10.1002/mp.70024
Vancouver
Urago Y, Sakama M, Sakata D, Katayose T, Chang W. Monte Carlo–calculated beam–quality and perturbation correction factors for helium–, carbon–, and neon–ion beams. Medical Physics. 2025. doi:10.1002/mp.70024
BibTeX
@article{Urago2025,
author = {Urago, Yuka and Sakama, Makoto and Sakata, Dousatsu and Katayose, Tetsurou and Chang, Weishan},
title = {Monte Carlo–calculated beam–quality and perturbation correction factors for helium–, carbon–, and neon–ion beams},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.70024}
}
EUVIMED