Small dose monitor based on silicon‐carbide diodes for FLASH radiotherapy

Ivan Lopez Paz iD, Celeste Fleta iD, Ángela Henao iD, Sophie Heinrich iD, Consuelo Guardiola iD

Medical Physics · 2026-03 · 53 (3)

DOI: 10.1002/mp.70354 ↗

At a glance

Journal article in Medical Physics, published online on 2026-02-27. Written by Ivan Lopez Paz and 4 co-authors. Institution: Institute of Microelectronics of Barcelona, IMB-CNM (CSIC), Barcelona, Spain. Major MEDLINE topics: Silicon Compounds, Carbon Compounds, Inorganic, Radiometry, Radiotherapy. Cited 1 times so far (Crossref); 57 references. The full text is freely available (open access, CC BY 4.0). Funded by: LaCaixa Health foundation DosiFLASH, UHDPulse, European Union's Horizon 2020 research and innovation program.

AbstractSubjectsBibliographic detailsCiteRelated studies

Abstract (original language)

Background

The FLASH biological effect in radiotherapy has been observed to appear at ultra‐high dose rates UHDR (40 Gy/s), where the accurate dosimetry at such high rates is still a challenge.

Purpose

A new 4 4 array of SiC‐based detectors (1 mm diameter, 2.2 mm pitch) is proposed for dosimetry in UHDR, as well as the feasibility of a position sensitive technology demonstrator covering 7 7 placed on a movable micro‐stage to cover larger surfaces.

Methods

In the ElectronFlash LINAC at the Institute Curie, two silicon carbide prototypes (a 2.2 mm diameter single diode and a 4 4‐array of 1 mm diameter with a pitch of 2.2 mm), biased at 0 V, are exposed to a 0.5–5 pulsed electron beam of 7 MeV alongside a flashDiamond PTW as reference dosimeter to characterize their response, time structure and position response.

Results

A linearity better than 3.5% is observed up to 10 Gy per pulse of the single diode device only limited by the reference dosimetry. The pulse structure measured is consistent with the reference beam current transformer installed in the LINAC, allowing for instantaneous pulse discrimination at UHDR and its verification in the measurement point. Moreover, results demonstrate the viability of using SiC arrays to quantify the dose per pulse in a 70 50 area with a granularity of 1 2.2 , paving the way to larger arrays and thus toward potential 2D dose monitoring.

Conclusions

The possibility of a position sensitive SiC dose monitor for UHDR is demonstrated, as the technology demonstrator has been proven to maintain good linearity up to at least 10 Gy per pulse, with a time resolution enough to observe microsecond pulses and position sensitive readout.

Subjects

MeSH terms (MEDLINE)

Silicon CompoundsCarbon Compounds, InorganicRadiometryRadiotherapyRadiotherapy DosageEquipment DesignParticle Accelerators

Author keywords

dosimeterssilicon carbideultra‐high dose rate

Substances

silicon carbideSilicon CompoundsCarbon Compounds, Inorganic

Authors and institutions

  • Institute of Microelectronics of Barcelona, IMB-CNM (CSIC), Barcelona, Spain.
  • Institut Curie, Inserm U1021-CNRS UMR 3347, University Paris-Saclay, PSL Research University, Centre Universitaire, Orsay, France.

ORCID profiles: Ivan Lopez Paz, Celeste Fleta, Ángela Henao, Sophie Heinrich, Consuelo Guardiola

Funding

  • LaCaixa Health foundation DosiFLASH
  • UHDPulse
  • European Union's Horizon 2020 research and innovation program
  • TWAC: EIC Pathfinder Open 2021
  • CSIC PIE Incorporación CT
  • NEWDOSI
  • Agencia Estatal de Investigación

Cite

APA

Lopez Paz, I., Fleta, C., Henao, Á., Heinrich, S., & Guardiola, C. (2026). Small dose monitor based on silicon‐carbide diodes for FLASH radiotherapy. Medical Physics, 53(3). https://doi.org/10.1002/mp.70354

Vancouver

Lopez Paz I, Fleta C, Henao Á, Heinrich S, Guardiola C. Small dose monitor based on silicon‐carbide diodes for FLASH radiotherapy. Medical Physics. 2026;53(3). doi:10.1002/mp.70354

BibTeX
@article{LopezPaz2026,
  author = {Lopez Paz, Ivan and Fleta, Celeste and Henao, Ángela and Heinrich, Sophie and Guardiola, Consuelo},
  title = {Small dose monitor based on silicon‐carbide diodes for FLASH radiotherapy},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {3},
  doi = {10.1002/mp.70354}
}
Start for free →