Field output correction factors and perturbation factor analysis for novel SunSILICON and SunSILICON P silicon diode detectors
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70632 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Andreas A Schönfeld und 13 weiteren Autor:innen. Beteiligte Einrichtung: Sun Nuclear Corp., Melbourne, Florida, USA. Hauptthemen laut MEDLINE: Particle Accelerators, Monte Carlo Method, Silicon, Photons. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Field output factor measurements of clinical linear accelerators require detector‐specific corrections, due to the introduction of fluence perturbation effects by non‐water‐equivalent components of applied detectors.
Purpose
This study aims to determine field output correction factors for novel shielded and unshielded silicon diode detectors in high energy photon fields. Special emphasis is placed on understanding the influence of perturbation factors and evaluating the suitability of these detectors across a wide range of field sizes.
Methods
Field output correction factors for the silicon diodes SunSILICON and SunSILICON P were determined through experimental measurements conducted at four distinct sites. Monte Carlo‐based models of the diode detectors were developed to calculate field output correction factors using the EGSnrc code system. In addition, these silicon diode models enabled a detailed analysis of perturbation factors for field sizes from 0.6 cm up to 40 cm.
Results
Experimentally determined and Monte Carlo calculated field output correction factors are in good agreement for the detectors investigated. The perturbation factor analysis demonstrated a strong field size dependence of the fluence perturbation factor for silicon.
Conclusion
This is the first study to systematically characterize the SunSILICON diode detector family using both simulation and measurement. The findings confirm that both shielded and unshielded designs are suitable for clinical dosimetry in photon beams, requiring only minor corrections. The shielding of the SunSILICON P diode enables accurate field output factor measurements across a broad range of field sizes, establishing its utility in modern radiotherapy applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Sun Nuclear Corp., Melbourne, Florida, USA.
- CDT-West-Centrum für Diagnostik und Therapie, Cologne, Germany.
- Ärztepartnerschaft Radiologie Vechta, Vechta, Germany.
- Universität Bremen, Bremen, Germany.
- Mirion Technologies, Olen, Belgium.
- MyMichigan Health, Midland, Michigan, USA.
Zitieren
APA
Schönfeld, A. A., Alissa, M., Towle, D., Murray, A., Schönfeld, A. B., Anders, T. J., Alquist, E., Evrard, O., Rosenwald, C., Hildreth, J., Flatten, V., Wessing, G., Habib, C., & Czarnecki, D. (2026). Field output correction factors and perturbation factor analysis for novel SunSILICON and SunSILICON P silicon diode detectors. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70632
Vancouver
Schönfeld AA, Alissa M, Towle D, Murray A, Schönfeld AB, Anders TJ, et al. Field output correction factors and perturbation factor analysis for novel SunSILICON and SunSILICON P silicon diode detectors. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70632
BibTeX
@article{Schnfeld2026,
author = {Schönfeld, Andreas A and Alissa, Mohamad and Towle, David and Murray, Andy and Schönfeld, Ann-Britt and Anders, Teresa J and Alquist, Erik and Evrard, Olivier and Rosenwald, Chuck and Hildreth, Jeff and Flatten, Veronika and Wessing, Gerhard and Habib, Charbel and Czarnecki, Damian},
title = {Field output correction factors and perturbation factor analysis for novel SunSILICON and SunSILICON P silicon diode detectors},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70632}
}
EUVIMED