Investigating the use of a solid water and copper electron filter for reducing skin dose in an inline MRI‐linac
Medical Physics · 2025
DOI: 10.1002/mp.70166 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Madiha Tai und 5 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Copper, Skin, Water. Die Arbeit nennt 44 Literaturangaben. Gefördert von: Federal Government, Australian Research Council Discovery.
Abstract (Originalsprache)
Background
Skin dose measurements for the inline MRI‐linac show a highly concentrated skin dose hot spot at the central axis of the beam. This is because the electron contamination is focused in the presence of the inline magnetic field, causing a significant entry surface (or skin) dose hot spot. The future clinical treatment will employ a 2 cm water electron filter to reduce electron contamination from the X‐ray beam; however, this approach affects the overall dose distribution, and the skin dose remains significantly high.
Purpose
To evaluate and compare the effectiveness of a copper electron filter (CuEF) and a water electron filter (WEF) in reducing the skin dose on the inline MRI‐linac, using Monte Carlo simulations and experimental measurements.
Methods
The magnetic field maps, designed using a 3D magnetic field model for a 1 T MRI‐linac in COMSOL Multiphysics, were subsequently integrated into the Geant4 Monte Carlo model of the Australian MRI‐linac. The Geant4 geometry also included the WEF, CuEF and a 30 30 30 cm 3 water phantom. Simulations were performed for field sizes 2.4 2.4 cm 2 , 7.2 7.2 cm 2 and 12.4 12.4 cm 2 . To compare the efficiency of a 2 cm water block in mitigating the electron contamination, simulations were performed with WEF positioned above the phantom surface. Central axis percentage depth dose and surface skin dose values at 70 depth were measured using high‐resolution scoring voxels. Additional simulations were performed by placing CuEF at varying distances from the MRI isocenter. The simulation results were further validated by the experimental measurements performed using the MO Skin ™ detector by incorporating the physical prototypes of the CuEF and WEF in the pre‐clinical setup of MRI‐linac.
Results
In the presence of a magnetic field, the skin dose values were measured to be 122.5% (2.4 2.4 cm 2 ), 194.8% (7.2 7.2 cm 2 ), and 260.4% (12.4 12.4 cm 2 ) of the dose at d max (1.5 cm depth) for near 0 T measurements. Adding the WEF 30 cm above the phantom surface reduces these skin dose values; however, the hot spot's intensity remains substantial, and the skin dose is higher than d max . The CuEF was found to be more effective in reducing the skin dose hot spots across all three field sizes. Positioning the CuEF 10 cm above the phantom surface reduced the skin dose values by about 25%, 53%, and 64% for the field size 2.4 2.4 cm 2 , 7.2 7.2 cm 2 , and 12.4 12.4 cm 2 , respectively.
Conclusions
Incorporating the CuEF above the phantom surface in the inline MRI‐linac prototype notably reduces the intensity of the skin dose hot spot, proving to be a more effective solution than a standard 2 cm water block across all field sizes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia.
- GenesisCare, St Vincent's Hospital, Darlinghurst, NSW, Australia.
- Department of Medical Physics, Liverpool and Macarthur Cancer Therapy Centre, Liverpool, NSW, Australia.
- Ingham Institute for Applied Medical Research, Liverpool, NSW, Australia.
- South Western Sydney Clinical School, University of New South Wales, Liverpool, NSW, Australia.
- GenesisCare, Frankston Private Hospital, Frankston, VIC, Australia.
Förderung
- Federal Government
- Australian Research Council Discovery
Zitieren
APA
Tai, M., Begg, J., Patterson, E., Metcalfe, P. E., Rosenfeld, A., & Oborn, B. M. (2025). Investigating the use of a solid water and copper electron filter for reducing skin dose in an inline MRI‐linac. Medical Physics. https://doi.org/10.1002/mp.70166
Vancouver
Tai M, Begg J, Patterson E, Metcalfe PE, Rosenfeld A, Oborn BM. Investigating the use of a solid water and copper electron filter for reducing skin dose in an inline MRI‐linac. Medical Physics. 2025. doi:10.1002/mp.70166
BibTeX
@article{Tai2025,
author = {Tai, Madiha and Begg, Jarrad and Patterson, Elizabeth and Metcalfe, Peter E and Rosenfeld, Anatoly and Oborn, Bradley M},
title = {Investigating the use of a solid water and copper electron filter for reducing skin dose in an inline MRI‐linac},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.70166}
}
EUVIMED