EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Dose perturbations at tissue interfaces during parallel linac‐MR treatments: The “Lateral Scatter Electron Return Effect” (LS‐ERE)

Stephen Steciw, B Gino Fallone, Eugene Yip

Medical Physics · 2024

DOI: 10.1002/mp.17363 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Stephen Steciw und 2 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Division, Department of Oncology, University of Alberta, Edmonton, Alberta, Canada. Hauptthemen laut MEDLINE: Particle Accelerators, Magnetic Resonance Imaging, Electrons, Monte Carlo Method. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Magnetic resonance (MR) imaging devices have been integrated with medical linear accelerators (linac) in radiation therapy. Both perpendicular linac‐MR (LMR‐B⊥) and parallel (LMR‐B∥) systems exist, where due to the MR's magnetic field dose can be perturbed in the patient. Dose perturbations from the electron return effect (ERE) and electron streaming effects (ESEs) are present in LMR‐B⊥ systems, where a dose collimating effect has been observed in LMR‐B∥ systems .

Purpose

To report on an asymmetric dose perturbation which is present at the interface between two different materials during treatment in parallel linac‐MR (LMR‐B∥) systems. To the best of our knowledge, these asymmetric dose effects, “Lateral Scattered Electron Return Effect” (LS‐ERE) have not been previously reported.

Methods

BEAMnrc and EGSnrc Monte Carlo (MC) radiation transport codes were used with the EEMF macro to emulate a 6 FFF beam from the 0.5‐T Alberta linac‐MR (LMR). Simulations were performed at 0.5 and 1.5 T in several different phantom material–interface combinations and field sizes including from modulated MLC‐like fields. MC simulations quantified LS‐ERE in patient CT datasets for the head, breast, and lung. LS‐ERE cancellation techniques were investigated. LS‐ERE asymmetries were quantified by subtracting an antiparallel dose from the parallel dose, dividing by two and normalizing to the global 0‐T maximum dose. GafChromic film measurements were made in the 0.5‐T Alberta LMR‐B∥ system using solid water at the water–air interface to validate MC simulations. ERE was simulated for an emulated LMR‐B⊥ system and compared to LMR‐B∥ dose perturbations.

Results

LS‐ERE is mostly independent of field size for fields >1 × 1 cm 2 . For 5 × 5‐cm 2 fields at 0.5T/1.5T, LS‐ERE asymmetries are ≤±6.9%/6.9% at bone–air and ≤±9.0%/7.0% at tissue–air for nonair doses, and ≤±4.1%/5.5% at tissue–lung interfaces. LS‐ERE increases as the density gradient increases, where the magnitude and extent of LS‐ERE are reduced as field strength increases. For a single 5 × 5‐cm 2 field at 0.5T/1.5T, the LS‐ERE asymmetry is ≤±10.2%/8.5% at the tissue–air sinus interface for head, ≤±4.2%/5.3% at the spine–lung interface for the lung, and ≤±5.7%/4.9% at the skin–air interface for a breast tangent plan at 0.5T/1.5T. POP fields mostly remove LS‐ERE asymmetries, with magnetic field reversal during treatment being the most effective method. Skin dose was investigated and compared to 0‐T treatments for 0.5T/1.5T LMR‐B∥ single field breast and head treatments. Including all dosimetric magnetic field perturbations, a 21%/24% and 22%/22% increase in skin dose to head and breast, respectively, was observed, of which LS‐ERE is responsible for approximately 30% of the total. Measured LS‐ERE asymmetries and dose enhancements at the water–air interface using GafChromic film were in excellent agreement with MC simulations. ERE in 1.5‐T LMR‐B⊥ systems are on average 5.5 times larger than total dose perturbations at 0.5 T in LMR‐B∥ systems.

Conclusion

LS‐ERE is present at the interface between materials and awareness of LS‐ERE is crucial for proper TPS evaluation for LMR‐B∥ treatments, especially in areas where large tissue density gradients exist.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Particle AcceleratorsMagnetic Resonance ImagingElectronsMonte Carlo MethodRadiotherapy DosageHumansScattering, RadiationRadiation DosagePhantoms, Imaging

Stichwörter der Autor:innen

EREdose perturbationparallel

Autor:innen und Institutionen

  • Medical Physics Division, Department of Oncology, University of Alberta, Edmonton, Alberta, Canada.
  • Department of Medical Physics, Cross Cancer Institute, Edmonton, Alberta, Canada.

Zitieren

APA

Steciw, S., Fallone, B. G., & Yip, E. (2024). Dose perturbations at tissue interfaces during parallel linac‐MR treatments: The “Lateral Scatter Electron Return Effect” (LS‐ERE). Medical Physics. https://doi.org/10.1002/mp.17363

Vancouver

Steciw S, Fallone BG, Yip E. Dose perturbations at tissue interfaces during parallel linac‐MR treatments: The “Lateral Scatter Electron Return Effect” (LS‐ERE). Medical Physics. 2024. doi:10.1002/mp.17363

BibTeX
@article{Steciw2024,
  author = {Steciw, Stephen and Fallone, B Gino and Yip, Eugene},
  title = {Dose perturbations at tissue interfaces during parallel linac‐MR treatments: The “Lateral Scatter Electron Return Effect” (LS‐ERE)},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17363}
}
Kostenlos starten →