Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14370 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Elizabeth Patterson und 5 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Medical and Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia. Hauptthemen laut MEDLINE: Phantoms, Imaging, Particle Accelerators, Electrons, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To evaluate the accuracy of different dosimeters and the treatment planning system (TPS) for assessing the skin dose due to the electron streaming effect (ESE) on a 1.5 T magnetic resonance (MR)‐linac.
Method
Skin dose due to the ESE on an MR‐linac (Unity, Elekta) was investigated using a solid water phantom rotated 45° in the x‐y plane (IEC61217) and centered at the isocenter. The phantom was irradiated with 1 × 1, 3 × 3, 5 × 5, 10 × 10, and 22 × 22 cm 2 fields, gantry at 90°. Out‐of‐field doses (OFDs) deposited by electron streams generated at the entry and exit surface of the angled phantom were measured on the surface of solid water slabs placed ±20.0 cm from the isocenter along the x‐direction. A high‐resolution MO Skin ™ detector served as a benchmark due to its shallower depth of measurement that matches the International Commission on Radiological Protection (ICRP) recommended depth for skin dose assessment (0.07 mm). MO Skin ™ doses were compared to EBT3 film, OSLDs, a diamond detector, and the TPS where the experimental setup was modeled using two separate calculation parameters settings: a 0.1 cm dose grid with 0.2% statistical uncertainty (0.1 cm, 0.2%) and a 0.2 cm dose grid with 3.0% statistical uncertainty (0.2 cm, 3.0%).
Results
OSLD, film, the 0.1 cm, 0.2%, and 0.2 cm, 3.0% TPS ESE doses, underestimated skin doses measured by the MO Skin ™ by as much as –75.3%, –7.0%, –24.7%, and –41.9%, respectively. Film results were most similar to MO Skin ™ skin dose measurements.
Conclusions
These results show that electron streams can deposit significant doses outside the primary field and that dosimeter choice and TPS calculation settings greatly influence the reported readings. Due to the steep dose gradient of the ESE, EBT3 film remains the choice for accurate skin dose assessment in this challenging environment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Centre for Medical and Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
- College of Science and Engineering, James Cook University, Townsville, Queensland, Australia.
- Townsville Cancer Centre, Townsville Hospital and Health Service, Townsville, Queensland, Australia.
- Illawarra Health Medical Research Institute, University of Wollongong, Wollongong, New South Wales, Australia.
- Department of Radiation Oncology, St George Cancer Care Centre, Wollongong, New South Wales, Australia.
- Institute of Radiooncology- OncoRay, Helmholtz-Zentrum Dresden-Rossendorf, Radiooncology, Dresden, Germany.
ORCID-Profile: Elizabeth Patterson, Marcus Powers, Peter E Metcalfe, Dean Cutajar, Bradley M Oborn, John A Baines
Zitieren
APA
Patterson, E., Powers, M., Metcalfe, P. E., Cutajar, D., Oborn, B. M., & Baines, J. A. (2024). Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14370
Vancouver
Patterson E, Powers M, Metcalfe PE, Cutajar D, Oborn BM, Baines JA. Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14370
BibTeX
@article{Patterson2024,
author = {Patterson, Elizabeth and Powers, Marcus and Metcalfe, Peter E and Cutajar, Dean and Oborn, Bradley M and Baines, John A},
title = {Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14370}
}
EUVIMED