Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac

Elizabeth Patterson iD, Marcus Powers iD, Peter E Metcalfe iD, Dean Cutajar iD, Bradley M Oborn iD, John A Baines iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14370 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Elizabeth Patterson und 5 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Medical and Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia. Hauptthemen laut MEDLINE: Phantoms, Imaging, Particle Accelerators, Electrons, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the accuracy of different dosimeters and the treatment planning system (TPS) for assessing the skin dose due to the electron streaming effect (ESE) on a 1.5 T magnetic resonance (MR)‐linac.

Method

Skin dose due to the ESE on an MR‐linac (Unity, Elekta) was investigated using a solid water phantom rotated 45° in the x‐y plane (IEC61217) and centered at the isocenter. The phantom was irradiated with 1 × 1, 3 × 3, 5 × 5, 10 × 10, and 22 × 22 cm 2 fields, gantry at 90°. Out‐of‐field doses (OFDs) deposited by electron streams generated at the entry and exit surface of the angled phantom were measured on the surface of solid water slabs placed ±20.0 cm from the isocenter along the x‐direction. A high‐resolution MO Skin ™ detector served as a benchmark due to its shallower depth of measurement that matches the International Commission on Radiological Protection (ICRP) recommended depth for skin dose assessment (0.07 mm). MO Skin ™ doses were compared to EBT3 film, OSLDs, a diamond detector, and the TPS where the experimental setup was modeled using two separate calculation parameters settings: a 0.1 cm dose grid with 0.2% statistical uncertainty (0.1 cm, 0.2%) and a 0.2 cm dose grid with 3.0% statistical uncertainty (0.2 cm, 3.0%).

Results

OSLD, film, the 0.1 cm, 0.2%, and 0.2 cm, 3.0% TPS ESE doses, underestimated skin doses measured by the MO Skin ™ by as much as –75.3%, –7.0%, –24.7%, and –41.9%, respectively. Film results were most similar to MO Skin ™ skin dose measurements.

Conclusions

These results show that electron streams can deposit significant doses outside the primary field and that dosimeter choice and TPS calculation settings greatly influence the reported readings. Due to the steep dose gradient of the ESE, EBT3 film remains the choice for accurate skin dose assessment in this challenging environment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingParticle AcceleratorsElectronsRadiotherapy Planning, Computer-AssistedRadiotherapy DosageMagnetic Resonance ImagingHumansRadiotherapy, Intensity-ModulatedSkinMonte Carlo Method

Stichwörter der Autor:innen

Elekta UnityMRgRTMR‐linacMonaco Unityelectron streaming effectout‐of‐field doseskin dose

Autor:innen und Institutionen

  • Centre for Medical and Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • College of Science and Engineering, James Cook University, Townsville, Queensland, Australia.
  • Townsville Cancer Centre, Townsville Hospital and Health Service, Townsville, Queensland, Australia.
  • Illawarra Health Medical Research Institute, University of Wollongong, Wollongong, New South Wales, Australia.
  • Department of Radiation Oncology, St George Cancer Care Centre, Wollongong, New South Wales, Australia.
  • Institute of Radiooncology- OncoRay, Helmholtz-Zentrum Dresden-Rossendorf, Radiooncology, Dresden, Germany.

ORCID-Profile: Elizabeth Patterson, Marcus Powers, Peter E Metcalfe, Dean Cutajar, Bradley M Oborn, John A Baines

Zitieren

APA

Patterson, E., Powers, M., Metcalfe, P. E., Cutajar, D., Oborn, B. M., & Baines, J. A. (2024). Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14370

Vancouver

Patterson E, Powers M, Metcalfe PE, Cutajar D, Oborn BM, Baines JA. Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14370

BibTeX
@article{Patterson2024,
  author = {Patterson, Elizabeth and Powers, Marcus and Metcalfe, Peter E and Cutajar, Dean and Oborn, Bradley M and Baines, John A},
  title = {Electron streaming dose measurements and calculations on a 1.5 T MR‐Linac},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14370}
}
Kostenlos starten →