Design and characterization of a novel scintillator array for UHDR PBS proton therapy surface dosimetry

Roman Vasyltsiv, Joseph Harms, Megan Clark, David J Gladstone, Brian W Pogue, Rongxiao Zhang, Petr Bruza

Medical Physics · 2025

DOI: 10.1002/mp.17922 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Roman Vasyltsiv und 6 weiteren Autor:innen. Beteiligte Einrichtung: Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA. Hauptthemen laut MEDLINE: Proton Therapy, Radiometry, Scintillation Counting. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: V Foundation for Cancer Research, NIH HHS, ROI.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Ultrahigh dose rate (UHDR) proton therapy has shown promise in normal tissue sparing by enhancing the therapeutic ratio through a method termed the FLASH effect. As in all radiotherapy, accurate in vivo dosimetry is crucial for quality assurance of safe and efficient treatment delivery. However, this remains a challenge for UHDR as existing dosimetry systems lack the spatial and temporal resolution required to verify dose and dose rate in complex anatomical regions, especially for pencil beam scanning (PBS) proton therapy.

Purpose

This study aims to develop and evaluate a novel 3D surface dosimetry method for UHDR PBS proton therapy using high‐speed imaging of a scintillator array, coupled with stereovision to provide real‐time, high‐resolution surface dose monitoring during treatment. The spatial, temporal, and dosimetric components of the proposed system are validated via imaging of a custom QA phantom and are compared against a gafchromic film reading of the same field delivered onto a flat surface.

Methods

A freely deformable multielement scintillator array was designed with a single element pitch of 7.5 mm and interelement gap of 0.5 mm. Scintillation linearity with dose was evaluated along with the variation in scintillator response with increasing imaging and irradiation angles. Water‐equivalent thickness (WET) testing was conducted to evaluate beam attenuation at two energy levels. Scintillation emission in response to dose delivery was imaged at 1000 Hz using a high frame rate camera (BeamSite Ultra, DoseOptics LLC) and the array position was monitored via a 2‐camera stereovision system. Imaging system setup was validated using a custom 3D QA phantom to assess spatial accuracy and guide systematic setup correction. Stereovision properties of each array element were used to guide angular emission correction, and geometric transformation to beams‐eye‐view (BEV). Kernel‐based residual spot fitting was applied to derive cumulative dose maps which were then compared to the flat film dose profile of a 5 × 5 cm UHDR PBS delivery using 3%/2 mm gamma analysis. PBS and maximum dose rate maps were also calculated.

Results

System setup achieved an average localization error of 0.62 mm, surpassing the typical 1+ mm threshold used in clinical practice. Intensity correction based on angular information was applied and yielded a cumulative spot dose uncertainty of ∼1% (5.428 mGy). The processed dose map was compared to film via gamma analysis with 3%/2 mm criteria and showed a 99.9% passing rate, indicating high agreement between the planned and measured dose profiles. The WET of the scintillator array was measured to be 1.1 mm, minimizing its impact on dose distribution.

Conclusion

The novel scintillator array system provides accurate, real‐time surface dose monitoring with high spatial and temporal resolution, making it a promising tool for in vivo dosimetry in UHDR proton therapy. Future work will focus on optimizing the system and expanding its application to other modalities, such as photon and electron therapy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyRadiometryScintillation CountingRadiotherapy DosageEquipment DesignPhantoms, Imaging

Autor:innen und Institutionen

  • Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA.
  • Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama, USA.
  • Department of Radiation Oncology, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin, USA.
  • Department of Radiation Oncology, University of Missouri, Columbia, Missouri, USA.

Förderung

  • V Foundation for Cancer Research
  • NIH HHS
  • ROI
  • NCI NIH HHS
  • NIBIB NIH HHS
  • National Institutes of Health
  • Research Opportunities Initiative, University of North Carolina

Zitieren

APA

Vasyltsiv, R., Harms, J., Clark, M., Gladstone, D. J., Pogue, B. W., Zhang, R., & Bruza, P. (2025). Design and characterization of a novel scintillator array for UHDR PBS proton therapy surface dosimetry. Medical Physics. https://doi.org/10.1002/mp.17922

Vancouver

Vasyltsiv R, Harms J, Clark M, Gladstone DJ, Pogue BW, Zhang R, et al. Design and characterization of a novel scintillator array for UHDR PBS proton therapy surface dosimetry. Medical Physics. 2025. doi:10.1002/mp.17922

BibTeX
@article{Vasyltsiv2025,
  author = {Vasyltsiv, Roman and Harms, Joseph and Clark, Megan and Gladstone, David J and Pogue, Brian W and Zhang, Rongxiao and Bruza, Petr},
  title = {Design and characterization of a novel scintillator array for UHDR PBS proton therapy surface dosimetry},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17922}
}
Kostenlos starten →