EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Development of in situ dosimetry for FLASH proton radiotherapy via organic scintillating fibers

Codey Olson, Jacob Strong, Adam Paxton, Geoff Nelson, Edward Cazalas iD

British Journal of Radiology · 2026

DOI: 10.1093/bjr/tqag152 ↗

Kurz gefasst

Fachartikel in der Zeitschrift British Journal of Radiology, 2026. Verfasst von Codey Olson und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Civil and Environmental Engineering, University of Utah Nuclear Engineering Program, Salt Lake City, UT 84112, United States. Hauptthemen laut MEDLINE: Proton Therapy, Radiometry, Scintillation Counting. Die Arbeit nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This work presents the design, development, and initial proof-of-concept testing of a dosimetry device intended for FLASH radiotherapy.

Methods

The system incorporates organic scintillating fibers coupled to a silicon photomultiplier (SiPM) and an FPGA-based readout, with fiber geometry optimized to minimize perturbations to a clinical proton beam, as confirmed through measurements at the Huntsman Cancer Institute (HCI).

Results

Tests using a portable X-ray unit demonstrated a strong correlation between integrated charge and delivered dose (up to ∼35 mrem s−1) and showed that the detector can respond on nanosecond timescales, enabling rapid safety interlock triggering with minimal latency.

Conclusion

Preliminary proton irradiations at HCI further indicated that the device can resolve individual beam pulses, a capability valuable for both FLASH research and conventional radiotherapy. Advances in knowledge These results highlight the need for continued refinement of the experimental setup, readout electronics, and detector performance to fully realize real-time, in situ dose and pulse monitoring.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyRadiometryScintillation CountingEquipment DesignRadiotherapy DosageHumans

Stichwörter der Autor:innen

FLASHGEANT4dosimetryprotonradiotherapy

Autor:innen und Institutionen

  • Department of Civil and Environmental Engineering, University of Utah Nuclear Engineering Program, Salt Lake City, UT 84112, United States.
  • Department of Radiation Oncology, Huntsman Cancer Institute, Salt Lake City, UT 84112, United States.

ORCID-Profile: Edward Cazalas

Zitieren

APA

Olson, C., Strong, J., Paxton, A., Nelson, G., & Cazalas, E. (2026). Development of in situ dosimetry for FLASH proton radiotherapy via organic scintillating fibers. British Journal of Radiology. https://doi.org/10.1093/bjr/tqag152

Vancouver

Olson C, Strong J, Paxton A, Nelson G, Cazalas E. Development of in situ dosimetry for FLASH proton radiotherapy via organic scintillating fibers. British Journal of Radiology. 2026. doi:10.1093/bjr/tqag152

BibTeX
@article{Olson2026,
  author = {Olson, Codey and Strong, Jacob and Paxton, Adam and Nelson, Geoff and Cazalas, Edward},
  title = {Development of in situ dosimetry for FLASH proton radiotherapy via organic scintillating fibers},
  journal = {British Journal of Radiology},
  year = {2026},
  doi = {10.1093/bjr/tqag152}
}