Improving decomposition image quality in dual‐energy chest radiography using two‐dimensional crisscrossed anti‐scatter grid

Duhee Jeon, Younghwan Lim, Hyesun Yang, Myeongkyu Park, Kyong-Woo Kim, Hyosung Cho

Medical Physics · 2025

DOI: 10.1002/mp.17819 ↗

At a glance

Journal article in Medical Physics, 2025. Written by Duhee Jeon and 5 co-authors. Institution: Department of Radiation Convergence Engineering, Yonsei University, Wonju, Republic of Korea. Major MEDLINE topics: Radiography, Thoracic, Scattering, Radiation, Image Processing, Computer-Assisted. Cited 5 times so far (Crossref); 53 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: Korea Medical Device Development Fund.

PovzetekSubjectsBibliografski podatkiCiteRelated studies

Abstract (original language)

Background

Chest radiography is a widely used medical imaging modality for diagnosing chest‐related diseases. However, anatomical structure overlap hinders accurate lesion detection. While the dual‐energy x‐ray imaging technique addresses this issue by separating soft‐tissue and bone images from an original chest radiograph, scattered radiation remains a significant challenge in decomposition image quality.

Purpose

This work aims to conduct dual‐energy material decomposition (DEMD) in chest radiography using a two‐dimensional (2D) crisscrossed anti‐scatter grid to improve decomposition image quality by effectively removing scattered radiation.

Methods

A 2D graphite‐interspaced grid with a strip density of N = 1.724 lines/mm and grid ratio r = 6:1 was fabricated using a high‐precision sawing process. The grid characteristics were evaluated using the IEC standard fixture. A 2D‐grid‐based DEMD process, which involves the acquisition of low‐ and high‐kV radiographs with a 2D grid, generation of a pairwise decomposition function using a calibration wedge phantom, and decomposition of soft‐tissue and bone images using the decomposition function, was implemented, followed by software‐based grid artifact reduction. Experiments were conducted on a commercially available chest phantom using an x‐ray imaging system operating at two tube voltages of 70 and 120 kVp. The decomposition image quality of the proposed DEMD and conventional dual‐energy subtraction methods was compared for the cases of no grid, software‐based scatter correction, 1D grid ( N = 8.475 lines/mm and r = 12:1), and 2D grid.

Results

The 2D grid demonstrated superior scatter radiation removal ability with scatter radiation transmission of 6.34% and grid selectivity of 9.67, representing a 2.6‐fold decrease and a 2.7‐fold improvement over the 1D grid, respectively. Compared to other competitive methods, the 2D‐grid‐based DEMD method considerably improved decomposition image quality, with improved lung structure visibility in selective soft‐tissue images.

Conclusions

The proposed DEMD method yielded high‐quality dual‐energy chest radiographs by effectively removing scattered radiation, demonstrating significant potential for improving lesion detection in clinical practice.

Subjects

MeSH terms (MEDLINE)

Radiography, ThoracicScattering, RadiationImage Processing, Computer-AssistedPhantoms, ImagingHumansArtifacts

Author keywords

chest radiographydual‐energy material decompositiondual‐energy subtractiontwo‐dimensional anti‐scatter grid

Authors and institutions

  • Department of Radiation Convergence Engineering, Yonsei University, Wonju, Republic of Korea.
  • R&D Center, JPI Healthcare Co., Ltd., Seoul, Republic of Korea.

Funding

  • Korea Medical Device Development Fund

Cite

APA

Jeon, D., Lim, Y., Yang, H., Park, M., Kim, K. W., & Cho, H. (2025). Improving decomposition image quality in dual‐energy chest radiography using two‐dimensional crisscrossed anti‐scatter grid. Medical Physics. https://doi.org/10.1002/mp.17819

Vancouver

Jeon D, Lim Y, Yang H, Park M, Kim KW, Cho H. Improving decomposition image quality in dual‐energy chest radiography using two‐dimensional crisscrossed anti‐scatter grid. Medical Physics. 2025. doi:10.1002/mp.17819

BibTeX
@article{Jeon2025,
  author = {Jeon, Duhee and Lim, Younghwan and Yang, Hyesun and Park, Myeongkyu and Kim, Kyong-Woo and Cho, Hyosung},
  title = {Improving decomposition image quality in dual‐energy chest radiography using two‐dimensional crisscrossed anti‐scatter grid},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17819}
}
Start for free →