A novel scatter correction method for energy-resolving photon-counting detector based CBCT imaging
Journal of X-Ray Science and Technology · 2025
DOI: 10.1177/08953996251351618 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of X-Ray Science and Technology, 2025. Verfasst von Xin Zhang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences…. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Image Processing, Computer-Assisted. Die Arbeit nennt 32 Literaturangaben. Gefördert von: Key Technologies Research and Development Program, National Natural Science Foundation of China, Shenzhen Science and Technology Innovation Program.
Abstract (Originalsprache)
Background
To generate high-quality CT images for an energy-resolving photon-counting detector (PCD) based cone beam CT (CBCT) system, it is essential to mitigate the scatter shading artifacts.
Objective
The aim of this study is to explore the capability of an energy-modulated scatter correction method, named e-Grid, in removing the scatter shading artifacts in energy-resolving PCD CBCT imaging.
Methods
In the e-Grid method, a linear approximation is assumed between the high-energy primary/scatter signals and the low-energy primary/scatter signals acquired from the two energy windows of a PCD. Calibration experiments were conducted to determine the parameters used in the aforementioned signal model. Physical validation experiments with head and abdominal phantoms were performed on a PCD CBCT imaging benchtop system.
Results
It was found that the e-Grid method could significantly eliminate scatter cupping artifacts in both low-energy and high-energy PCD CBCT imaging for objects with varying dimensions. Quantitatively, results demonstrated that the e-Grid method reduced scatter artifacts by more than 70% in both low-energy and high-energy PCD CBCT images.
Conclusions
In this study, it is demonstrated that the e-Grid scatter correction method has great potential for reducing scatter shading artifacts in energy-resolving PCD CBCT imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.
- University of Chinese Academy of Sciences, Beijing, China.
- Paul C Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.
- State Key Laboratory of Biomedical Imaging Science and System, Shenzhen, China.
- National Innovation Center for Advanced Medical Devices, Shenzhen, China.
- Research Center for Medical Artificial Intelligence, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.
ORCID-Profile: Yongshuai Ge
Förderung
- Key Technologies Research and Development Program
- National Natural Science Foundation of China
- Shenzhen Science and Technology Innovation Program
- Youth Innovation Promotion Association of the Chinese Academy of Sciences
Zitieren
APA
Zhang, X., Wang, H., Tan, Y., Zhu, J., Zheng, H., Liang, D., & Ge, Y. (2025). A novel scatter correction method for energy-resolving photon-counting detector based CBCT imaging. Journal of X-Ray Science and Technology. https://doi.org/10.1177/08953996251351618
Vancouver
Zhang X, Wang H, Tan Y, Zhu J, Zheng H, Liang D, et al. A novel scatter correction method for energy-resolving photon-counting detector based CBCT imaging. Journal of X-Ray Science and Technology. 2025. doi:10.1177/08953996251351618
BibTeX
@article{Zhang2025,
author = {Zhang, Xin and Wang, Heran and Tan, Yuhang and Zhu, Jiongtao and Zheng, Hairong and Liang, Dong and Ge, Yongshuai},
title = {A novel scatter correction method for energy-resolving photon-counting detector based CBCT imaging},
journal = {Journal of X-Ray Science and Technology},
year = {2025},
doi = {10.1177/08953996251351618}
}
EUVIMED