A novel scatter correction method for energy-resolving photon-counting detector based CBCT imaging

Xin Zhang, Heran Wang, Yuhang Tan, Jiongtao Zhu, Hairong Zheng, Dong Liang, Yongshuai Ge iD

Journal of X-Ray Science and Technology · 2025

DOI: 10.1177/08953996251351618 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of X-Ray Science and Technology, 2025. Verfasst von Xin Zhang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences…. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Image Processing, Computer-Assisted. Die Arbeit nennt 32 Literaturangaben. Gefördert von: Key Technologies Research and Development Program, National Natural Science Foundation of China, Shenzhen Science and Technology Innovation Program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

To generate high-quality CT images for an energy-resolving photon-counting detector (PCD) based cone beam CT (CBCT) system, it is essential to mitigate the scatter shading artifacts.

Objective

The aim of this study is to explore the capability of an energy-modulated scatter correction method, named e-Grid, in removing the scatter shading artifacts in energy-resolving PCD CBCT imaging.

Methods

In the e-Grid method, a linear approximation is assumed between the high-energy primary/scatter signals and the low-energy primary/scatter signals acquired from the two energy windows of a PCD. Calibration experiments were conducted to determine the parameters used in the aforementioned signal model. Physical validation experiments with head and abdominal phantoms were performed on a PCD CBCT imaging benchtop system.

Results

It was found that the e-Grid method could significantly eliminate scatter cupping artifacts in both low-energy and high-energy PCD CBCT imaging for objects with varying dimensions. Quantitatively, results demonstrated that the e-Grid method reduced scatter artifacts by more than 70% in both low-energy and high-energy PCD CBCT images.

Conclusions

In this study, it is demonstrated that the e-Grid scatter correction method has great potential for reducing scatter shading artifacts in energy-resolving PCD CBCT imaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyImage Processing, Computer-AssistedPhantoms, ImagingPhotonsArtifactsScattering, RadiationHumansHeadAlgorithmsAbdomen

Stichwörter der Autor:innen

CBCT imagingphoton-counting detectorscatter

Autor:innen und Institutionen

  • Research Center for Advanced Detection Materials and Medical Imaging Devices, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.
  • University of Chinese Academy of Sciences, Beijing, China.
  • Paul C Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.
  • State Key Laboratory of Biomedical Imaging Science and System, Shenzhen, China.
  • National Innovation Center for Advanced Medical Devices, Shenzhen, China.
  • Research Center for Medical Artificial Intelligence, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China.

ORCID-Profile: Yongshuai Ge

Förderung

  • Key Technologies Research and Development Program
  • National Natural Science Foundation of China
  • Shenzhen Science and Technology Innovation Program
  • Youth Innovation Promotion Association of the Chinese Academy of Sciences

Zitieren

APA

Zhang, X., Wang, H., Tan, Y., Zhu, J., Zheng, H., Liang, D., & Ge, Y. (2025). A novel scatter correction method for energy-resolving photon-counting detector based CBCT imaging. Journal of X-Ray Science and Technology. https://doi.org/10.1177/08953996251351618

Vancouver

Zhang X, Wang H, Tan Y, Zhu J, Zheng H, Liang D, et al. A novel scatter correction method for energy-resolving photon-counting detector based CBCT imaging. Journal of X-Ray Science and Technology. 2025. doi:10.1177/08953996251351618

BibTeX
@article{Zhang2025,
  author = {Zhang, Xin and Wang, Heran and Tan, Yuhang and Zhu, Jiongtao and Zheng, Hairong and Liang, Dong and Ge, Yongshuai},
  title = {A novel scatter correction method for energy-resolving photon-counting detector based CBCT imaging},
  journal = {Journal of X-Ray Science and Technology},
  year = {2025},
  doi = {10.1177/08953996251351618}
}
Kostenlos starten →