Rapid CNN‐based needle localization for automatic slice alignment in MR‐guided interventions using 3D undersampled radial white‐marker imaging

Jonas Frederik Faust, Axel Joachim Krafft, Daniel Polak, Peter Speier, Nicolas Gerhard Roland Behl, Nathan Ooms, Jesse Roll, Joshua Krieger, Mark Edward Ladd, Florian Maier

Medical Physics · 2024

DOI: 10.1002/mp.17376 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Jonas Frederik Faust and 9 co-authors. Institution: Faculty of Physics and Astronomy, Ruprecht-Karls-Universität Heidelberg, Heidelberg, Germany. Major MEDLINE topics: Needles, Imaging, Three-Dimensional, Neural Networks, Computer. Cited 1 times so far (Crossref); 29 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

In MR‐guided in‐bore percutaneous needle interventions, typically 2D interactive real‐time imaging is used for navigating the needle into the target. Misaligned 2D imaging planes can result in losing visibility of the needle in the 2D images, which impedes successful targeting. Necessary iterative manual slice adjustment can prolong interventional workflows. Therefore, rapid automatic alignment of the imaging planes with the needle would be preferable to improve such workflows.

Purpose

To investigate rapid 3D localization of needles in MR‐guided interventions via a convolutional neural network (CNN)‐based localization algorithm using an undersampled white‐marker contrast acquisition for the purpose of automatic imaging slice alignment.

Methods

A radial 3D rf‐spoiled gradient echo MR pulse sequence with white‐marker encoding was implemented and a CNN‐based localization algorithm was employed to extract position and orientation of an aspiration needle from the undersampled white‐marker images. The CNN was trained using porcine tissue phantoms (257 needle trajectories, four‐fold data augmentation, 90%/10% split into training and validation dataset). Achievable localization times and accuracy were evaluated retrospectively in an ex vivo study (109 needle trajectories) for a range of needle orientations between 78° and 90° relative to the B 0 field. A proof‐of‐concept in vivo experiment was performed in two porcine animal models and feasibility of automatic imaging slice alignment was evaluated retrospectively.

Results

Ex vivo needle localization was achieved with a median localization accuracy of 1.9 mm (distance needle tip to detected needle axis) and a median angular deviation of 2.6° for needle orientations between 86° and 90° to the B 0 field from fully sampled WM images (resolution of (4 mm) 3 , 6434 acquired radial k‐space spokes, acquisition time of 80.4 s) in a field‐of‐view of (256 mm) 3 . Localization accuracy decreased with increasing undersampling and needle trajectory increasingly aligned with B 0 . For needle orientations between 86° and 90° to the B 0 field, a highly accelerated acquisition of only 32 k‐space spokes (acquisition time of 0.4 s) yielded a median localization accuracy of 3.1 mm and a median angular deviation of 4.7°. For needle orientations between 78° and 82° to the B 0 field, a median accuracy and angular deviation of 3.5 mm and 6.8° could still be achieved with 64 sampled spokes (acquisition time of 0.8 s). In vivo, a localization accuracy of 1.4 mm and angular deviation of 3.4° was achieved sampling 32 k‐space spokes (acquisition time of 0.48 s) with the needle oriented at 87.7° to the B 0 field. For a needle oriented at 77.6° to the B 0 field, localization accuracy of 5.3 mm and angular deviation of 6.8° were still achieved sampling 128 k‐space spokes (acquisition time of 1.92 s), allowing for retrospective slice alignment.

Conclusion

The investigated approach enables passive biopsy needle localization in 3D. Acceleration of the localization to real‐time applicability is feasible for needle orientations approximately perpendicular to B 0 . The method can potentially facilitate MR‐guided needle interventions by enabling automatic imaging slice alignment with the needle.

Subjects

MeSH terms (MEDLINE)

NeedlesImaging, Three-DimensionalNeural Networks, ComputerSwineAnimalsAutomationMagnetic Resonance ImagingPhantoms, ImagingTime FactorsAlgorithms

Author keywords

device localizationinterventional MRIpercutaneous needle intervention

Authors and institutions

  • Faculty of Physics and Astronomy, Ruprecht-Karls-Universität Heidelberg, Heidelberg, Germany.
  • Siemens Healthineers AG, Erlangen, Germany.
  • Cook Advanced Technologies, West Lafayette, Indiana, USA.
  • School of Health Sciences, Purdue University, West Lafayette, Indiana, USA.
  • Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
  • Faculty of Medicine, Ruprecht-Karls-Universität Heidelberg, Heidelberg, Germany.

Cite

APA

Faust, J. F., Krafft, A. J., Polak, D., Speier, P., Behl, N. G. R., Ooms, N., Roll, J., Krieger, J., Ladd, M. E., & Maier, F. (2024). Rapid CNN‐based needle localization for automatic slice alignment in MR‐guided interventions using 3D undersampled radial white‐marker imaging. Medical Physics. https://doi.org/10.1002/mp.17376

Vancouver

Faust JF, Krafft AJ, Polak D, Speier P, Behl NGR, Ooms N, et al. Rapid CNN‐based needle localization for automatic slice alignment in MR‐guided interventions using 3D undersampled radial white‐marker imaging. Medical Physics. 2024. doi:10.1002/mp.17376

BibTeX
@article{Faust2024,
  author = {Faust, Jonas Frederik and Krafft, Axel Joachim and Polak, Daniel and Speier, Peter and Behl, Nicolas Gerhard Roland and Ooms, Nathan and Roll, Jesse and Krieger, Joshua and Ladd, Mark Edward and Maier, Florian},
  title = {Rapid CNN‐based needle localization for automatic slice alignment in MR‐guided interventions using 3D undersampled radial white‐marker imaging},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17376}
}
Start for free →