A workflow for predicting radiofrequency‐induced heating around bilateral deep brain stimulation electrodes in MRI

Nur Izzati Huda Zulkarnain, Alireza Sadeghi-Tarakameh, Jeromy Thotland, Noam Harel, Yigitcan Eryaman

Medical Physics · 2024

DOI: 10.1002/mp.16913 ↗

At a glance

Journal article in Medical Physics, 2024. Written by Nur Izzati Huda Zulkarnain and 4 co-authors. Institution: Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, Minnesota, USA. Major MEDLINE topics: Heating, Deep Brain Stimulation. Cited 7 times so far (Crossref); 62 references. The full text is freely available (open access, CC BY 4.0). Funded by: NIBIB NIH HHS, NCI NIH HHS, NINDS NIH HHS.

SamenvattingSubjectsBibliografische gegevensCiteRelated studies

Abstract (original language)

Background

Heating around deep brain stimulation (DBS) in magnetic resonance imaging (MRI) occurs when the time‐varying electromagnetic (EM) fields induce currents in the electrodes which can generate heat and potentially cause tissue damage. Predicting the heating around the electrode contacts is important to ensure the safety of patients with DBS implants undergoing an MRI scan. We previously proposed a workflow to predict heating around DBS contacts and introduced a parameter, equivalent transimpedance, that is independent of electrode trajectories, termination, and radiofrequency (RF) excitations. The workflow performance was validated in a unilateral DBS system.

Purpose

To predict RF heating around the contacts of bilateral (DBS) electrodes during an MRI scan in an anthropomorphic head phantom.

Methods

Bilateral electrodes were fixed in a skull phantom filled with hydroxyethyl cellulose (HEC) gel. The electrode shafts were suspended extracranially, in a head and torso phantom filled with the same gel material. The current induced on the electrode shaft was experimentally measured using an MR‐based technique 3 cm above the tip. A transimpedance value determined in a previous offline calibration was used to scale the shaft current and calculate the contact voltage. The voltage was assigned as a boundary condition on the electrical contacts of the electrode in a quasi‐static (EM) simulation. The resulting specific absorption rate (SAR) distribution became the input for a transient thermal simulation and was used to predict the heating around the contacts. RF heating experiments were performed for eight different lead trajectories using circularly polarized (CP) excitation and two linear excitations for one trajectory. The measured temperatures for all experiments were compared with the simulated temperatures and the root‐mean‐squared errors (RMSE) were calculated.

Results

The RF heating around the contacts of both bilateral electrodes was predicted with ≤ 0.29°C of RMSE for 20 heating scenarios.

Conclusion

The workflow successfully predicted the heating for different bilateral DBS trajectories and excitation patterns in an anthropomorphic head phantom.

Subjects

MeSH terms (MEDLINE)

HeatingDeep Brain StimulationHumansWorkflowMagnetic Resonance ImagingElectrodesRadio WavesPhantoms, ImagingElectrodes, Implanted

Author keywords

MRIdeep brain stimulation safetyradiofrequency heatingtemperature prediction

Authors and institutions

  • Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, Minnesota, USA.

Funding

  • NIBIB NIH HHS
  • NCI NIH HHS
  • NINDS NIH HHS
  • National Institute of Biomedical Imaging and Bioengineering
  • National Institute of Neurological Disorders and Stroke

Cite

APA

Zulkarnain, N. I. H., Sadeghi-Tarakameh, A., Thotland, J., Harel, N., & Eryaman, Y. (2024). A workflow for predicting radiofrequency‐induced heating around bilateral deep brain stimulation electrodes in MRI. Medical Physics. https://doi.org/10.1002/mp.16913

Vancouver

Zulkarnain NIH, Sadeghi-Tarakameh A, Thotland J, Harel N, Eryaman Y. A workflow for predicting radiofrequency‐induced heating around bilateral deep brain stimulation electrodes in MRI. Medical Physics. 2024. doi:10.1002/mp.16913

BibTeX
@article{Zulkarnain2024,
  author = {Zulkarnain, Nur Izzati Huda and Sadeghi-Tarakameh, Alireza and Thotland, Jeromy and Harel, Noam and Eryaman, Yigitcan},
  title = {A workflow for predicting radiofrequency‐induced heating around bilateral deep brain stimulation electrodes in MRI},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16913}
}
Start for free →