EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

MRI of Implantation Sites Using Parallel Transmission of an Optimized Radiofrequency Excitation Vector

Mostafa Berangi iD, Andre Kuehne iD, Helmar Waiczies, Thoralf Niendorf iD

Tomography · 2023

DOI: 10.3390/tomography9020049 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Tomography, 2023. Verfasst von Mostafa Berangi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Berlin Ultrahigh Field Facility, Max-Delbrück-Center for Molecular Medicine in the Helmholtz Association, 13125 Berlin, Germany. Hauptthemen laut MEDLINE: Radio Waves, Magnetic Resonance Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 61 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union’s Horizon 2020 research and innovation program under the Marie Skłodowska-Curie, European Research Council (ERC) under the European Union Horizon 2020 research and innovation program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Postoperative care of orthopedic implants is aided by imaging to assess the healing process and the implant status. MRI of implantation sites might be compromised by radiofrequency (RF) heating and RF transmission field (B1+) inhomogeneities induced by electrically conducting implants. This study examines the applicability of safe and B1+-distortion-free MRI of implantation sites using optimized parallel RF field transmission (pTx) based on a multi-objective genetic algorithm (GA). Electromagnetic field simulations were performed for eight eight-channel RF array configurations (f = 297.2 MHz), and the most efficient array was manufactured for phantom experiments at 7.0 T. Circular polarization (CP) and orthogonal projection (OP) algorithms were applied for benchmarking the GA-based shimming. B1+ mapping and MR thermometry and imaging were performed using phantoms mimicking muscle containing conductive implants. The local SAR10g of the entire phantom in GA was 12% and 43.8% less than the CP and OP, respectively. Experimental temperature mapping using the CP yielded ΔT = 2.5–3.0 K, whereas the GA induced no extra heating. GA-based shimming eliminated B1+ artefacts at implantation sites and enabled uniform gradient-echo MRI. To conclude, parallel RF transmission with GA-based excitation vectors provides a technical foundation en route to safe and B1+-distortion-free MRI of implantation sites.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radio WavesMagnetic Resonance ImagingPhantoms, ImagingAlgorithmsHeating

Stichwörter der Autor:innen

MRIRF arrayimplantparallel transmissionradiofrequencysafetytransmission field shimming

Autor:innen und Institutionen

  • Berlin Ultrahigh Field Facility, Max-Delbrück-Center for Molecular Medicine in the Helmholtz Association, 13125 Berlin, Germany.
  • Charité-Universitätsmedizin Berlin, 10117 Berlin, Germany.
  • MRI.TOOLS GmbH, 13125 Berlin, Germany.

ORCID-Profile: Mostafa Berangi, Andre Kuehne, Thoralf Niendorf

Förderung

  • European Union’s Horizon 2020 research and innovation program under the Marie Skłodowska-Curie
  • European Research Council (ERC) under the European Union Horizon 2020 research and innovation program

Zitieren

APA

Berangi, M., Kuehne, A., Waiczies, H., & Niendorf, T. (2023). MRI of Implantation Sites Using Parallel Transmission of an Optimized Radiofrequency Excitation Vector. Tomography. https://doi.org/10.3390/tomography9020049

Vancouver

Berangi M, Kuehne A, Waiczies H, Niendorf T. MRI of Implantation Sites Using Parallel Transmission of an Optimized Radiofrequency Excitation Vector. Tomography. 2023. doi:10.3390/tomography9020049

BibTeX
@article{Berangi2023,
  author = {Berangi, Mostafa and Kuehne, Andre and Waiczies, Helmar and Niendorf, Thoralf},
  title = {MRI of Implantation Sites Using Parallel Transmission of an Optimized Radiofrequency Excitation Vector},
  journal = {Tomography},
  year = {2023},
  doi = {10.3390/tomography9020049}
}
Kostenlos starten →