Comparative study of radiofrequency heating in deep brain stimulation devices during MRI at 1.5 T and 0.55 T: Challenging the assumption of safety at low field strengths

Bhumi Bhusal iD, Pia Panravi Sanpitak, Fuchang Jiang iD, Jacob Richardson, Nicole Seiberlich, Joshua M Rosenow, Behzad Elahi, Laleh Golestanirad iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30515 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Bhumi Bhusal und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Northwestern University, Chicago, Illinois, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Deep Brain Stimulation, Radio Waves. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIBIB NIH HHS, National Institutes of Health (NIH).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Low‐field MRI has been assumed to be implant‐friendly based on limited studies. However, radiofrequency (RF)–induced heating due to an implant is a complex resonance phenomenon, highly dependent on the implant's configurations and applied RF frequencies. This study aims to evaluate the RF heating of deep brain stimulation (DBS) implants during MRI at low field strengths compared with higher‐field 1.5T MRI.

Methods

A commercial DBS implant was used in full system as well as lead‐only configurations to evaluate and compare RF heating during MRI at 0.55 T and 1.5 T. The transfer function of the device was measured and validated at each of the frequencies, which was then used for prediction of RF heating with realistic DBS configurations at head, chest, and abdomen imaging landmarks.

Results

For the lead‐only case, the RF heating due to the DBS was substantially smaller during imaging at 0.55 T compared with that at 1.5 T. However, for the full DBS system (longer implant), the RF heating at 0.55 T was comparable to, and for some cases even higher than, that at 1.5 T, reaching a level that poses risk of tissue damage in patients.

Conclusions

Although RF heating generally tends to be lower at low‐field MRI, the case with longer implanted leads demands extra caution, due to the higher possibility of matching resonant condition at low‐field‐strength frequencies. Thus, specific risk evaluation for each implant and configuration is required rather than assuming that lower field strength imaging is safer.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingDeep Brain StimulationRadio WavesHumansHot TemperatureBrainPhantoms, Imaging

Stichwörter der Autor:innen

MRI safetyRF heatingdeep brain stimulationlow‐field MRImedical implantnumerical modeling

Autor:innen und Institutionen

  • Department of Radiology, Northwestern University, Chicago, Illinois, USA.
  • Department of Biomedical Engineering, Northwestern University, Evanston, Illinois, USA.
  • Department of Radiology, University of Michigan, Ann Arbor, Michigan, USA.
  • Department of Neurological Surgery, Northwestern University, Chicago, Illinois, USA.
  • Department of Neurology and Neurological Surgery, Loyola University Medical Center, Maywood, Illinois, USA.

ORCID-Profile: Bhumi Bhusal, Fuchang Jiang, Laleh Golestanirad

Förderung

  • NIBIB NIH HHS
  • National Institutes of Health (NIH)

Zitieren

APA

Bhusal, B., Sanpitak, P. P., Jiang, F., Richardson, J., Seiberlich, N., Rosenow, J. M., Elahi, B., & Golestanirad, L. (2025). Comparative study of radiofrequency heating in deep brain stimulation devices during MRI at 1.5 T and 0.55 T: Challenging the assumption of safety at low field strengths. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30515

Vancouver

Bhusal B, Sanpitak PP, Jiang F, Richardson J, Seiberlich N, Rosenow JM, et al. Comparative study of radiofrequency heating in deep brain stimulation devices during MRI at 1.5 T and 0.55 T: Challenging the assumption of safety at low field strengths. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30515

BibTeX
@article{Bhusal2025,
  author = {Bhusal, Bhumi and Sanpitak, Pia Panravi and Jiang, Fuchang and Richardson, Jacob and Seiberlich, Nicole and Rosenow, Joshua M and Elahi, Behzad and Golestanirad, Laleh},
  title = {Comparative study of radiofrequency heating in deep brain stimulation devices during MRI at 1.5 T and 0.55 T: Challenging the assumption of safety at low field strengths},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30515}
}
Kostenlos starten →