RF ‐induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T

Md Zahidul Islam, Ran Guo, Mir Khadiza Akter, Jianfeng Zheng iD, Wolfgang Kainz, Stuart Long, Ji Chen iD

Magnetic Resonance in Medicine · 2025

DOI: 10.1002/mrm.30389 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Md Zahidul Islam und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Radio Waves, Equipment and Supplies. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 51 Literaturangaben. Gefördert von: Division of Industrial Innovation and Partnerships, National Science Foundation, Directorate for Technology, Innovation and Partnerships.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To address the issue of RF‐induced heating for partially in and partially out (PIPO) medical devices during 1.5 T MRI scans by proposing a method of minimizing the external portion.

Methods

A method of tightly winding the external segment of the PIPO device is proposed to minimize the overall device effective reception length during MRI scans to mitigate the RF‐induced heating. Two commercially available PIPO medical devices and simplified solid wires were used to demonstrate the concept. RF heating results are compared between typical and minimized‐length trajectories under the American Society for Testing and Materials (ASTM) testing procedure. In addition, 16 scaled and validated device models were used in conjuncture with human body numerical simulations within three virtual human models to estimate clinically relevant heating.

Results

The wound segments in PIPO devices functioned as a lumped element rather than a receiving antenna, reducing induced energy/heating as compared to the original PIPO devices under typical straight or loop configurations. Minimizing the lead's external portion can reduce the RF‐induced heating by significant factors for all studied cases during ASTM phantom measurements and in human body simulations.

Conclusion

Our findings show a significant reduction in RF heating by minimizing the external segment, thereby enhancing patient safety during 1.5 T MRI procedures. Although limited to four devices at 1.5 T across two applications, the extent of heating reduction may vary for others. Nonetheless, tightly winding the external segment of PIPO electrodes holds promise for improving device safety under MRI.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingRadio WavesEquipment and SuppliesHumansEquipment DesignHot TemperatureComputer SimulationPhantoms, ImagingHeating

Stichwörter der Autor:innen

MR safetymagnetic resonance imaging (MRI)partially in and partially out (PIPO) medical implantsradio frequency (RF) heating mitigationtrajectory modificationtransfer function (TF)

Autor:innen und Institutionen

  • Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA.
  • High Performance Computing for MRI Safety, Jasper, Georgia, USA.

ORCID-Profile: Jianfeng Zheng, Ji Chen

Förderung

  • Division of Industrial Innovation and Partnerships
  • National Science Foundation
  • Directorate for Technology, Innovation and Partnerships

Zitieren

APA

Islam, M. Z., Guo, R., Akter, M. K., Zheng, J., Kainz, W., Long, S., & Chen, J. (2025). RF ‐induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30389

Vancouver

Islam MZ, Guo R, Akter MK, Zheng J, Kainz W, Long S, et al. RF ‐induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30389

BibTeX
@article{Islam2025,
  author = {Islam, Md Zahidul and Guo, Ran and Akter, Mir Khadiza and Zheng, Jianfeng and Kainz, Wolfgang and Long, Stuart and Chen, Ji},
  title = {RF ‐induced heating reduction by minimizing the external portion of the partially in and partially out medical devices under MRI at 1.5 T},
  journal = {Magnetic Resonance in Medicine},
  year = {2025},
  doi = {10.1002/mrm.30389}
}
Kostenlos starten →