Evaluation of a two‐dimensional Moire‐free antiscatter grid for cone‐beam computed tomography
Medical Physics · 2023-06 · 50 (6) · 3435-3444
DOI: 10.1002/mp.16243 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 17.02.2023. Verfasst von Jongkuk Kim und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Grid Research and Development, JPI Healthcare, Co., Ltd., Ansan-si, Gyeonggi-do, Republic of Korea. Hauptthemen laut MEDLINE: Software, Algorithms. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Ministry of Trade, Industry and Energy (Korea), JPI Healthcare Co., Ltd., Korea Medical Device Development Fund.
Abstract (Originalsprache)
Background
Scatter radiation is a traditional problem in cone‐beam computed tomography (CBCT). Accordingly, numerous methods have been researched for scatter reduction in terms of software‐ or hardware‐based solutions. The concept of a two‐dimensional antiscatter grid (2D‐ASG) has been shown to provide a solution to the scatter reduction in CBCT. However, the use of an ASG makes it challenging to use in clinical CBCT systems because it causes Moire artifacts of the image.
Purpose
We have developed a Moire‐free 2D‐ASG that was designed to solve the Moire artifact and the scatter radiation problems. We provide the experimental results pertaining to the image quality measurements from the evaluation of the 2D‐ASG compared with those obtained from a one‐dimensional antiscatter grid (1D‐ASG) and no antiscatter grid (No‐ASG) to demonstrate the quantitative extent of the improvements.
Methods
The 2D‐ASG, fabricated based on a sawing process with a graphite body, was prepared to evaluate image quality improvements. Projection images for Pro‐CT MK II phantom were acquired using the CBCT testbed of sample rotation type and reconstructed by Feldkamp–Davis–Kress (FDK) algorithm without any filters. We measured the contrast‐to‐noise ratio (CNR) and uniformity index for the cupping artifacts of the 2D‐ASG, 1D‐ASG, and No‐ASG cases.
Results
2D‐ASG considerably reduced the cupping artifacts owing to the scatter reduction compared with the 1D‐ASG and No‐ASG cases. The cupping artifacts were reduced by 85% in the 2D‐ASG compared with No‐ASG, while the cupping artifacts were reduced by 63% in 1D‐ASG compared with No‐ASG. The 2D‐ASG also yielded a CNR improvement. The average CNR improvements for eight insert materials were 47% in the 2D‐ASG compared with No‐ASG, while the CNR was improved by 36% in the 1D‐ASG compared with No‐ASG.
Conclusions
We demonstrated that the Moire‐free 2D‐ASG improved the image quality by removing scatter radiation in CBCT compared with 1D‐ASG and No‐ASG. We believe that the Moire‐free 2D‐ASG can become one of the effective ways to solve the scatter radiation problem in CBCT images because it provides usability and has the potential to have synergistic effects on other methods, such as bow‐tie filter and scatter correction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Grid Research and Development, JPI Healthcare, Co., Ltd., Ansan-si, Gyeonggi-do, Republic of Korea.
Förderung
- Ministry of Trade, Industry and Energy (Korea)
- JPI Healthcare Co., Ltd.
- Korea Medical Device Development Fund
Zitieren
APA
Kim, J., Kang, Y., Hwang, T., Park, M., & Chung, W. (2023). Evaluation of a two‐dimensional Moire‐free antiscatter grid for cone‐beam computed tomography. Medical Physics, 50(6), 3435-3444. https://doi.org/10.1002/mp.16243
Vancouver
Kim J, Kang Y, Hwang T, Park M, Chung W. Evaluation of a two‐dimensional Moire‐free antiscatter grid for cone‐beam computed tomography. Medical Physics. 2023;50(6):3435-3444. doi:10.1002/mp.16243
BibTeX
@article{Kim2023,
author = {Kim, Jongkuk and Kang, Yoonseok and Hwang, Taesoon and Park, Myungkyu and Chung, Woohyun},
title = {Evaluation of a two‐dimensional Moire‐free antiscatter grid for cone‐beam computed tomography},
journal = {Medical Physics},
year = {2023},
volume = {50},
number = {6},
pages = {3435--3444},
doi = {10.1002/mp.16243}
}
EUVIMED