Balloon Dilation Technique Influences Stent Shortening and Intentional Stent Fracture During Overexpansion of Drug‐Eluting Stents: A Bench‐Testing Study

Mudit Gupta iD, Ella B Leeth, Yuval Barak-Corren, Ari J Gartenberg iD, Elizabeth Silvestro, Michael L O'Byrne iD, Ryan Callahan

Catheterization and Cardiovascular Interventions · 2025

DOI: 10.1002/ccd.31428 ↗

At a glance

Vergleichsstudie in Catheterization and Cardiovascular Interventions, 2025. Written by Mudit Gupta and 6 co-authors. Institution: Department of Pediatrics, Division of Cardiology, The Children's Hospital of Philadelphia, University of Pennsylvania Perelman School of Medicine, Ph…. Major MEDLINE topics: Drug-Eluting Stents, Prosthesis Failure, Angioplasty, Balloon, Coronary. Cited 2 times so far (Crossref); 18 references. Funded by: This work was supported by NIH T32 HL0097915., NHLBI NIH HHS.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background and Aims Drug‐eluting stents (DES) are not designed for overexpansion to supra‐nominal diameters or intentional stent fracture (ISF). The optimal dilation technique to minimize stent shortening (SS) and achieve ISF to accommodate pediatric somatic growth has not been described.

Methods

Three sizes of two commonly used DES were implanted within a silicone model to simulate blood vessels. Each stent was serially dilated in 1 mm increments under fluoroscopy using three techniques: 2 cm length, semi‐compliant balloons (Technique 1), 2 cm, noncompliant balloons straddling the entire stent (Technique 2), or noncompliant balloons in an “inside‐out” manner (balloon shorter than stent or 2 cm balloon aligned with distal end of stent (no straddle); (Technique 3). Technique 1 crossed over to noncompliant balloons once stent “napkin‐ringed” (NR). Percent SS = (L nominal – L final )/L nominal * 100.

Results

Technique 1 resulted in the greatest SS (median 85%, IQR 82, 87) and universal napkin ringing. Technique 2 caused less SS (median 14%, IQR 7, 15), and Technique 3 caused the least SS (median 7%, IQR 3, 11). ISF was achieved in all, however, the inside–out technique caused ISF at smaller stent diameters (median 114% recommended postdilation limit vs. 122%–131%) and lower inflation pressures (median 15 vs. 28–29 ATM). ISF was achieved in Technique 1 after napkin ringing but required larger noncompliant balloons than other techniques (median 8.5 vs. 7 mm).

Conclusion

Inside–out dilation resulted in less SS and fracture at smaller diameters with lower inflation pressures. This technique may improve the ISF success rate of DES in pediatric patients.

Subjects

MeSH terms (MEDLINE)

Drug-Eluting StentsProsthesis FailureAngioplasty, Balloon, CoronaryProsthesis DesignMaterials TestingHumansModels, AnatomicModels, CardiovascularSilicones

Author keywords

congenital heart diseasecoronary stentdrug‐eluting stentinterventionpulmonary vein stenosis

Substances

Silicones

Authors and institutions

  • Department of Pediatrics, Division of Cardiology, The Children's Hospital of Philadelphia, University of Pennsylvania Perelman School of Medicine, Philadelphia, Pennsylvania, USA.
  • Department of Pediatrics, Division of Cardiology, Akron Children's Hospital, Northeast Ohio Medical University, Akron, Ohio, USA.
  • Department of Radiology, The Children's Hospital of Philadelphia, University of Pennsylvania Perelman School of Medicine, Philadelphia, Pennsylvania, USA.

ORCID profiles: Mudit Gupta, Ari J Gartenberg, Michael L O'Byrne

Funding

  • This work was supported by NIH T32 HL0097915.
  • NHLBI NIH HHS

Cite

APA

Gupta, M., Leeth, E. B., Barak-Corren, Y., Gartenberg, A. J., Silvestro, E., O'Byrne, M. L., & Callahan, R. (2025). Balloon Dilation Technique Influences Stent Shortening and Intentional Stent Fracture During Overexpansion of Drug‐Eluting Stents: A Bench‐Testing Study. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31428

Vancouver

Gupta M, Leeth EB, Barak-Corren Y, Gartenberg AJ, Silvestro E, O'Byrne ML, et al. Balloon Dilation Technique Influences Stent Shortening and Intentional Stent Fracture During Overexpansion of Drug‐Eluting Stents: A Bench‐Testing Study. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31428

BibTeX
@article{Gupta2025,
  author = {Gupta, Mudit and Leeth, Ella B and Barak-Corren, Yuval and Gartenberg, Ari J and Silvestro, Elizabeth and O'Byrne, Michael L and Callahan, Ryan},
  title = {Balloon Dilation Technique Influences Stent Shortening and Intentional Stent Fracture During Overexpansion of Drug‐Eluting Stents: A Bench‐Testing Study},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2025},
  doi = {10.1002/ccd.31428}
}
Start for free →