Impact of Different Valve-in-Valve Positions on Functional Results of the New Generation of Balloon-Expandable Transcatheter Heart Valve
The Thoracic and Cardiovascular Surgeon · 01.06.2025 · 73 (04) · 288-295
DOI: 10.1055/s-0044-1787701 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift The Thoracic and Cardiovascular Surgeon, erschienen am 01.06.2025. Verfasst von Najla Sadat und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac and Thoracic Vascular Surgery, University Medical Center Schleswig Holstein Lübeck Campus, Lübeck, Schleswig-Holstein, Germany. Hauptthemen laut MEDLINE: Heart Valve Prosthesis, Transcatheter Aortic Valve Replacement, Aortic Valve, Balloon Valvuloplasty. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 19 Literaturangaben.
Abstract (Originalsprache)
Objectives
Very precise positioning of the transcatheter heart valve (THV) inside the degenerated SAV is a crucial factor for valve-in-valve (ViV) procedure to achieve optimal hemodynamic results. Therefore, our study aimed to investigate the impact of implantation depth on functional results after ViV procedures in a standardized in vitro setting.
Methods
THV (SAPIEN 3 Ultra 23-mm size) and three SAV models (Magna Ease, Trifecta, and Hancock II—all 21-mm size) were tested at different circulatory conditions in five different positions of the THV (2–6 mm) inside the SAV. Mean pressure gradient (MPG), effective orifice area (EOA), geometric orifice area (GOAmax), and pinwheeling index (PWImean) were analyzed.
Results
EOA and MPG of the THV did not differ significantly regarding the position inside the Magna Ease and the Hancock II (p > 0.05). However, EOA differed significantly, depending on the position of the THV inside Trifecta (2 vs. 5 mm; p = 0.021 and 2 vs. 6 mm; p < 0.001). The THV presented the highest EOA (2.047 cm2) and the lowest MPG (5.387 mm Hg) inside the Magna Ease, whereas the lowest EOA (1.335 cm2) and the highest MPG (11.876 mm Hg) were shown inside the Hancock II. Additionally, the highest GOAmax and the lowest PWImean of the THV were noticed inside the Magna Ease. The THV showed lower GOAmax and higher PWImean inside the Trifecta when placed in a deeper position.
Conclusion
Deep implantation of the SAPIEN 3 Ultra inside the Trifecta correlates with impaired functional results. In contrast, the implantation position of the SAPIEN 3 Ultra inside the Magna Ease and the Hancock II did not have a significant effect on functional results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Cardiac and Thoracic Vascular Surgery, University Medical Center Schleswig Holstein Lübeck Campus, Lübeck, Schleswig-Holstein, Germany.
- Department of Cardio-Thoracic and Vascular Surgery, University Hospital Tübingen, Tübingen, Baden-Württemberg, Germany.
Zitieren
APA
Sadat, N., Scharfschwerdt, M., Reichert, S., Fujita, B., & Ensminger, S. (2025). Impact of Different Valve-in-Valve Positions on Functional Results of the New Generation of Balloon-Expandable Transcatheter Heart Valve. The Thoracic and Cardiovascular Surgeon, 73(04), 288-295. https://doi.org/10.1055/s-0044-1787701
Vancouver
Sadat N, Scharfschwerdt M, Reichert S, Fujita B, Ensminger S. Impact of Different Valve-in-Valve Positions on Functional Results of the New Generation of Balloon-Expandable Transcatheter Heart Valve. The Thoracic and Cardiovascular Surgeon. 2025;73(04):288-295. doi:10.1055/s-0044-1787701
BibTeX
@article{Sadat2025,
author = {Sadat, Najla and Scharfschwerdt, Michael and Reichert, Stefan and Fujita, Buntaro and Ensminger, Stephan},
title = {Impact of Different Valve-in-Valve Positions on Functional Results of the New Generation of Balloon-Expandable Transcatheter Heart Valve},
journal = {The Thoracic and Cardiovascular Surgeon},
year = {2025},
volume = {73},
number = {04},
pages = {288--295},
doi = {10.1055/s-0044-1787701}
}
EUVIMED