Retrospective analysis of machine service logs and EPID‐based linac QA to investigate the effects of machine maintenance and calibration events on beam characteristics: A multi‐institutional study

John DeMarco, Michael Barnes iD, Hania Al-Hallaq iD, Seng Boh Lim iD, Lana Critchfield, Cory S Knill, Karl Farrey, Timothy Ritter, Nilendu Gupta, Richard A Popple iD, Justin Mikell, Jim Irrer und 2 weitere

Journal of Applied Clinical Medical Physics · 2026-11 · 27 (11)

DOI: 10.1002/acm2.70894 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 29.10.2026. Verfasst von John DeMarco und 13 weiteren Autor:innen. Beteiligte Einrichtung: Cedars-Sinai Department of Radiation Oncology, NT Lower-Level, Los Angeles, California, USA. Hauptthemen laut MEDLINE: Particle Accelerators, Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, Quality Control. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Varian, Varian Medical Systems.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Electronic Portal Imaging Devices (EPIDs) are widely used for LINAC quality assurance (QA) and underpin many manufacturer integrated quality control (MIQC) tools, yet the practical impact of routine linac service and calibration events on EPID‐derived beam characteristics remains under‐described. The Automated Quality Assurance (AQA) consortium has developed an independent, standardized EPID workflow for QA assessment that has been deployed across multiple TrueBeam linacs. We leveraged the AQA's framework and data to time‐align vendor service logs with automated EPID analyses and characterize event‐linked changes in photon beam output, symmetry, and collimator‐jaw positioning.

Purpose

To investigate how EPID‐derived beam performance metrics change following common real‐world service events, including beam output calibration, active beam steering, and collimator jaw position calibration.

Methods

We analyzed a multi‐institutional dataset acquired between 2018 and 2024 across nine TrueBeam LINACs using the AQA Phase‐2 EPID suite. EPID acquisitions were delivered in MV dosimetry mode and uploaded to a web portal for automated analysis with corresponding XML service logs to align service‐event timelines with EPID QA results. Service events were grouped into beam‐generation (dose calibration and active beam steering) and collimator (jaw position calibration) categories. The vendor XML service logs categorized events using labels such as “ANGLE R/T” or “POSITION servo” for beam‐steering adjustments and “GAIN‐OFFSET” for jaw position calibration. The EPID analysis included center‐dose constancy, axial and transverse beam symmetry, collimator‐centering correction, and X‐/Y‐jaw position. Measurement outliers were flagged using a density‐based spatial clustering of applications with noise (DBSCAN) algorithm applied separately to one‐dimensional EPID time series.

Results

Across seven linacs with documented 6 MV dose calibrations, EPID center‐dose constancy showed a consistent post‐calibration decrease relative to baseline (mean ± SD = −1.10% ± 0.26%; range: −0.73% to −1.54%). In a paired subset (two linacs; n = 12 calibration events), the difference between EPID‐derived and ADCL‐calibrated ion‐chamber percent dose changes (ΔEPID − ΔIC) had a mean bias of 0.09% ± 0.79% (range −1.49% to 1.23%). Applying DBSCAN to the change in center‐dose output identified outlier points that were associated with lower (negative) changes across all nine linacs. Following ANGLE R/T steering, EPID‐derived symmetry shifts were observed and were concordant with daily detector‐array readings in a representative case, while symmetry changes also occurred after non‐steering activities (e.g., carousel/ion‐chamber work), indicating multiple service‐linked pathways affecting profile constancy. DBSCAN applied to jaw‐position metrics flagged outliers exceeding routine variability (e.g., 0.44 mm vs 0.03 mm), including a documented 1.3 mm Y2 shift temporally linked to a jaw GAIN‐OFFSET calibration.

Conclusions

Integrating service logs with automated EPID analysis enabled event‐level interpretation of longitudinal QA measurements. EPID center‐dose changes following 6 MV dose calibrations were independently compared with ADCL‐traceable ion‐chamber measurements for a limited subset of two linacs, whereas the symmetry, jaw‐position, and DBSCAN findings represent EPID‐internal associations with documented service events and were not independently verified. Standardized EPID calibration records and prospectively timed pre‐ and post‐service measurements are needed to strengthen attribution and confirm the broader applicability of these findings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Particle AcceleratorsQuality Assurance, Health CareRadiotherapy Planning, Computer-AssistedQuality ControlCalibrationHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

EPIDlinear acceleratorquality assurance

Autor:innen und Institutionen

  • Cedars-Sinai Department of Radiation Oncology, NT Lower-Level, Los Angeles, California, USA.
  • Department of Radiation Oncology, Calvary Mater Hospital Newcastle, Waratah, New South Wales, Australia.
  • Department of Radiation Oncology, Emory University, Atlanta, Georgia, USA.
  • Memorial Sloan-Kettering Cancer Center, Radiation Dosimetry Core, New York, New York, USA.
  • Department of Radiation Oncology, University of Michigan, Wyoming, Michigan, USA.
  • Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan, USA.

ORCID-Profile: Michael Barnes, Hania Al-Hallaq, Seng Boh Lim, Richard A Popple, Jean M Moran

Förderung

  • Varian
  • Varian Medical Systems

Zitieren

APA

DeMarco, J., Barnes, M., Al-Hallaq, H., Lim, S. B., Critchfield, L., Knill, C. S., Farrey, K., Ritter, T., Gupta, N., Popple, R. A., Mikell, J., Irrer, J., Perez, M., & Moran, J. M. (2026). Retrospective analysis of machine service logs and EPID‐based linac QA to investigate the effects of machine maintenance and calibration events on beam characteristics: A multi‐institutional study. Journal of Applied Clinical Medical Physics, 27(11). https://doi.org/10.1002/acm2.70894

Vancouver

DeMarco J, Barnes M, Al-Hallaq H, Lim SB, Critchfield L, Knill CS, et al. Retrospective analysis of machine service logs and EPID‐based linac QA to investigate the effects of machine maintenance and calibration events on beam characteristics: A multi‐institutional study. Journal of Applied Clinical Medical Physics. 2026;27(11). doi:10.1002/acm2.70894

BibTeX
@article{DeMarco2026,
  author = {DeMarco, John and Barnes, Michael and Al-Hallaq, Hania and Lim, Seng Boh and Critchfield, Lana and Knill, Cory S and Farrey, Karl and Ritter, Timothy and Gupta, Nilendu and Popple, Richard A and Mikell, Justin and Irrer, Jim and Perez, Mario and Moran, Jean M},
  title = {Retrospective analysis of machine service logs and EPID‐based linac QA to investigate the effects of machine maintenance and calibration events on beam characteristics: A multi‐institutional study},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {11},
  doi = {10.1002/acm2.70894}
}
Kostenlos starten →