An X‐ray and γ‐ray combination strategy to exploit steep dose gradients for improved prostate cancer radiotherapy: A dosimetric comparison of Taichi Pro and Halcyon
Journal of Applied Clinical Medical Physics · 2026-03 · 27 (3)
DOI: 10.1002/acm2.70532 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 27.02.2026. Verfasst von Kuo Li und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology Physics & Technology, Cancer Hospital of Shandong First Medical University, Jinan, China. Hauptthemen laut MEDLINE: Prostatic Neoplasms, Radiotherapy Planning, Computer-Assisted, Organs at Risk, Radiotherapy, Intensity-Modulated. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Collaborative Academic Innovation Project of Shandong Cancer Hospital, Taishan Scholars Program of Shandong Province.
Abstract (Originalsprache)
Purpose
This study aimed to evaluate a novel technological platform, Taichi Pro—which integrates a 6 MV flattening filterfree linear accelerator with a 18 source Rotating Gamma System (RGS) to generate steep dose gradients via multisource focused γ‐rays and noncoplanar arcs—for precision radiotherapy in prostate cancer. The work provides evidence to support the clinical adoption of hybridmodality radiotherapy devices.
Methods
Fifteen prostate cancer patients were enrolled. For each patient, a dual‐modality Taichi Pro plan (RGS focused on the planning gross tumor volume (PGTV) + Linac covering the planning target volume (PTV)) and a Halcyon photon plan (three‐arc VMAT) were designed while maintaining clinically tolerable dose to organs at risk (OAR). Comparative assessments included planning target volume (PTV and PGTV) metrics (D 95% , D mean , homogeneity index HI, conformity index CI), OAR doses (rectum V 40 /V 60 /D 2cc , bladder V 40 , testis D 2cc , etc.) and delivery efficiency to evaluate the ability to escalate target dose while sparing adjacent OARs.
Results
All plans met institutional clinical constraints. Taichi Pro significantly increased PGTV D mean (79.48 Gy ± 1.75 Gy) compared to Halcyon (73.25 Gy ± 0.55 24 Gy, P < 0.001) and D max (122.74 Gy ± 8.69 Gy) compared to Halcyon (76.15 Gy ± 0.79 Gy, P < 0.001), albeit poorer homogeneity (HI: 0.50 ± 0.09 for Taichi Pro vs. 0.06 ± 0.01 for Halcyon, though within clinically acceptable limits). Taichi Pro significantly reduced rectum V 60 (3.97% ± 3.25% vs. Halcyon 7.46% ± 4.78%, P = 0.016), and D 2cc (61.61 Gy ± 5.01 Gy vs. Halcyon 65.29 Gy ± 4.52 Gy, P = 0.040). Taichi Pro also significantly reduced testis D 2cc (2.39 ± 1.99 Gy) compared to (3.17 Gy ± 1.40 Gy, P = 0.006). Halcyon demonstrated significantly shorter beam‐on time (1.81 ± 0.23 minutes vs. 5.05 ± 1.59 minutes for Taichi Pro, P < 0.001).
Conclusion
Utilizing the steep dose gradient characteristic of the RGS, the Taichi Pro dual‐modality system effectively achieved target dose escalation while simultaneously improving sparing of adjacent OARs. This approach holds the potential for enhancing patient treatment outcomes and quality of life.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology Physics & Technology, Cancer Hospital of Shandong First Medical University, Jinan, China.
- Our United Corp, Xi'an, China.
Förderung
- National Natural Science Foundation of China
- Collaborative Academic Innovation Project of Shandong Cancer Hospital
- Taishan Scholars Program of Shandong Province
Zitieren
APA
Li, K., Yin, Y., Zhu, T., Sun, W., Wang, S., & Li, Z. (2026). An X‐ray and γ‐ray combination strategy to exploit steep dose gradients for improved prostate cancer radiotherapy: A dosimetric comparison of Taichi Pro and Halcyon. Journal of Applied Clinical Medical Physics, 27(3). https://doi.org/10.1002/acm2.70532
Vancouver
Li K, Yin Y, Zhu T, Sun W, Wang S, Li Z. An X‐ray and γ‐ray combination strategy to exploit steep dose gradients for improved prostate cancer radiotherapy: A dosimetric comparison of Taichi Pro and Halcyon. Journal of Applied Clinical Medical Physics. 2026;27(3). doi:10.1002/acm2.70532
BibTeX
@article{Li2026,
author = {Li, Kuo and Yin, Yong and Zhu, Ting and Sun, Weipeng and Wang, Shihao and Li, Zhenjiang},
title = {An X‐ray and γ‐ray combination strategy to exploit steep dose gradients for improved prostate cancer radiotherapy: A dosimetric comparison of Taichi Pro and Halcyon},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {3},
doi = {10.1002/acm2.70532}
}