Technical Notes: Robustness of three‐dimensional treatment and imaging isocenter testing using a new gel dosimeter and kilovoltage CBCT

Riki Oshika, Hidenobu Tachibana, Kazuya Seki, Rie Tachibana, Shunsuke Moriya, Takeji Sakae

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14439 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Riki Oshika und 5 weiteren Autor:innen. Beteiligte Einrichtung: Degree Programs in Comprehensive Human Sciences, Graduate School of Comprehensive Human Sciences, University of Tsukuba, Ibaraki, Japan. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, Radiosurgery. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Triangle Products Co., Ltd..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Coincidence of the treatment and imaging isocenter coordinates is required to safely perform small‐margin treatments, such as stereotactic radiosurgery of multiple brain metastases. A comprehensive and direct methodology for verifying concordance of kilovoltage cone‐beam computed tomography (kV‐CBCT) and treatment coordinates using an x‐ray CT‐based polymer gel dosimeter (dGEL) and onboard kV‐CBCT was previously reported. Using this methodology, we tested the ability of a new commercially available x‐ray CT‐based polymer dGEL with a rapid response to provide efficient quality assurance (QA).

Purpose

The aim of this study was to evaluate the robustness of the three‐dimensional geometric QA methodology using dGEL.

Methods

The dGEL were commercially manufactured. The prescribed dose for each field was determined by visually identifying the 5, 10, and 20 Gy isodose lines. A linear accelerator was used to irradiate the gels with seven non‐coplanar beams. An in‐house analysis program was used to identify the beam axes and treatment isocenter in kV‐CBCT coordinates by processing the pre‐ and post‐irradiation CBCT images. The impact of the radiation dose on the test reproducibility was examined, and the detectability of an intentional geometric error was assessed.

Results

The treatment isocenter was within 0.4 mm of the imaging isocenter for all radiation doses. The residual error of the test with the intentional error was within 0.2 mm. The analysis and image quality variations for a single dGEL introduced displacement errors less than 0.3 mm.

Conclusions

The test assessed the coincidence of treatment and kV‐CBCT isocenter coordinates and detected errors with high robustness. Even for a 10 Gy dose, the test yielded results comparable with those obtained using higher radiation doses owing to the rapid response of the dGEL dosimeter.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyRadiotherapy DosageRadiotherapy Planning, Computer-AssistedRadiosurgeryHumansRadiotherapy, Intensity-ModulatedGelsPhantoms, ImagingQuality Assurance, Health CareParticle AcceleratorsImage Processing, Computer-AssistedBrain NeoplasmsRadiation DosimetersImaging, Three-DimensionalRadiometry

Stichwörter der Autor:innen

3D starshotX‐ray CT‐based gel dosimeterdGEL™gel dosimetryrobustness

Substanzen

Gels

Autor:innen und Institutionen

  • Degree Programs in Comprehensive Human Sciences, Graduate School of Comprehensive Human Sciences, University of Tsukuba, Ibaraki, Japan.
  • Radiation Safety and Quality Assurance division, National Cancer Center Hospital East, Chiba, Japan.
  • Triangle Products Co. Ltd., Chiba, Japan.
  • Faculty of Medicine, University of Tsukuba, Ibaraki, Japan.

Förderung

  • Triangle Products Co., Ltd.

Zitieren

APA

Oshika, R., Tachibana, H., Seki, K., Tachibana, R., Moriya, S., & Sakae, T. (2024). Technical Notes: Robustness of three‐dimensional treatment and imaging isocenter testing using a new gel dosimeter and kilovoltage CBCT. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14439

Vancouver

Oshika R, Tachibana H, Seki K, Tachibana R, Moriya S, Sakae T. Technical Notes: Robustness of three‐dimensional treatment and imaging isocenter testing using a new gel dosimeter and kilovoltage CBCT. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14439

BibTeX
@article{Oshika2024,
  author = {Oshika, Riki and Tachibana, Hidenobu and Seki, Kazuya and Tachibana, Rie and Moriya, Shunsuke and Sakae, Takeji},
  title = {Technical Notes: Robustness of three‐dimensional treatment and imaging isocenter testing using a new gel dosimeter and kilovoltage CBCT},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14439}
}
Kostenlos starten →