Evaluation and implementation of an independent Kilovoltage X‐ray‐based imaging platform for carbon ion radiotherapy
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70501 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Yixiao Guo und 7 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou, Gansu, China. Hauptthemen laut MEDLINE: Phantoms, Imaging, Heavy Ion Radiotherapy, Cone-Beam Computed Tomography, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Central guidance local science and technology development program, Science and Technology Project of Lanzhou City, Scientific Research Management Project of Health Commission of Gansu Municipal.
Abstract (Originalsprache)
Background
Image‐guided particle therapy (IGPT) has significantly advanced in recent years, particularly in the context of proton therapy. However, imaging solutions for carbon‐ion radiotherapy (C‐ion RT) remain limited.
Purpose
This study introduces sliding‐gantry cone‐beam computed tomography (CBCT) and dual‐panel digital radiography (DR) systems, both mechanically independent of carbon‐ion delivery nozzles. We aim to evaluate the image quality metrics and verify the positioning accuracy of the imaging systems.
Methods
Image quality was evaluated in terms of spatial resolution, low contrast resolution, image uniformity, and effective imaging area using a multi‐purpose imaging phantom, Catphan 700 phantom, and ImageJ software. The influences of planning computed tomography (CT) slice thicknesses (1–5 mm), radiation quality settings (90–130 kV), and registration algorithms (bony, grayscale, and fiducial marker registrations) on positioning accuracy were assessed using anthropomorphic head‐neck and thoracoabdominal phantom images. The clinical feasibility of both systems was validated in 22 enrolled patients.
Results
The CBCT exhibited a lower in‐plane spatial resolution (2.50 line pairs per millimeter (lp/mm)) than DR (2.80 lp/mm). Spatial resolution of the CBCT system was measured at 0.90 lp/mm using the CTP 714 module of the Catphan 700 phantom. Both systems achieved a low contrast resolution of 2.30%. DR provided superior image uniformity (1.12%–1.40%) compared with CBCT (2.20%). The effective imaging areas were comparable between the CBCT and DR systems (99.30%–99.50%). Positioning accuracy varied with planning CT slice thicknesses, radiation quality settings, and registration algorithms, showing mean translation displacements ranging from 0.01 to 0.48 mm. CBCT achieved inter‐fraction translational positioning errors within 2 mm in 42.3% (22/52) of fractions and rotational positioning errors within 2° in 80.8% (42/52) of fractions, and DR achieved 33.8% (24/71) and 73.2% (52/71), respectively.
Conclusion
The developed CBCT and DR systems achieved superior image quality and sub‐0.5 mm positioning accuracy. These findings support the clinical feasibility of integrating CBCT and DR imaging systems into the C‐ion RT workflow.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou, Gansu, China.
- University of Chinese Academy of Sciences, Beijing, China.
- Gansu Provincial Hospital, Lanzhou, Gansu, China.
- LANZHOU ION THERAPY CO., LTD, Lanzhou, Gansu, China.
- Shanghai Shinetone Medical Equipment Co., Ltd, Shanghai, China.
- Heavy Ion Medicine Center, Wuhan University & People's Hospital of Wuhan Economic and Technological Development Zone, Wuhan, Hubei, China.
ORCID-Profile: Yixiao Guo, Zhiqiang Liu, Zhiguo Xu
Förderung
- Central guidance local science and technology development program
- Science and Technology Project of Lanzhou City
- Scientific Research Management Project of Health Commission of Gansu Municipal
- Natural Science Foundation of Gansu province
Zitieren
APA
Guo, Y., Liu, Z., Zhu, Q., Cai, M., Cai, H., Liu, R., Zhang, Q., & Xu, Z. (2026). Evaluation and implementation of an independent Kilovoltage X‐ray‐based imaging platform for carbon ion radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70501
Vancouver
Guo Y, Liu Z, Zhu Q, Cai M, Cai H, Liu R, et al. Evaluation and implementation of an independent Kilovoltage X‐ray‐based imaging platform for carbon ion radiotherapy. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70501
BibTeX
@article{Guo2026,
author = {Guo, Yixiao and Liu, Zhiqiang and Zhu, Qingzhen and Cai, Ming and Cai, Hongyi and Liu, Ruifeng and Zhang, Qiuning and Xu, Zhiguo},
title = {Evaluation and implementation of an independent Kilovoltage X‐ray‐based imaging platform for carbon ion radiotherapy},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70501}
}
EUVIMED