Characterization of a prototype rapid kilovoltage x‐ray image guidance system designed for a linear accelerator radiation therapy unit
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70060 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Theodore Arsenault und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University Hospitals Seidman Cancer Center, Cleveland, Ohio, USA. Hauptthemen laut MEDLINE: Particle Accelerators, Cone-Beam Computed Tomography, Algorithms, Phantoms, Imaging. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This study evaluates a novel cone‐beam computed tomography (CBCT) imaging solution integrated onto a conventional C‐arm linear accelerator (linac) with increased gantry speed. The purpose is to assess the impact of improved imaging hardware and reconstruction algorithms on image quality.
Methods
Hypersight‐CBCT (HS‐CBCT) system was compared with the original system (OG‐CBCT) on a TrueBeam linac. Performance tests included mechanical, geometric, setup accuracy, and image quality assessment. Image quality metrics were evaluated using conventional CBCT reconstruction and advanced iterative reconstruction (iCBCT), fast iCBCT, and iCBCT with metal artifact reduction. Dosimetry measurements were acquired.
Results
The HS‐CBCT system acquired images in 24.0‐44.0s (half trajectory/full trajectory), faster than the OG‐CBCT system's acquisition time of 33.5‐60.5s (half trajectory/full trajectory). The HS‐CBCT system's faster gantry speed resulted in comparable image quality to the OG‐CBCT. It improved high‐contrast resolution, modulation transfer function, and low‐contrast visibility. The faster gantry speed also produced lower radiation doses. The system's uniformity and resolution also improved, particularly with full‐fan acquisition techniques.
Conclusion
The novel HS‐CBCT system on a conventional C‐arm linac exhibits superior imaging capabilities compared to the OG‐CBCT. Faster gantry speed, larger imaging area, and advanced reconstruction algorithms contribute to enhanced image quality and reduced dose. The study provides comprehensive insights into the new system's performance, serving as a benchmark for future linac installations and highlighting potential benefits in clinical applications. Further investigations are suggested for 4D acquisitions and long‐term machine performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University Hospitals Seidman Cancer Center, Cleveland, Ohio, USA.
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
ORCID-Profile: Theodore Arsenault, Kenneth Gregg
Zitieren
APA
Arsenault, T., Arrue, J., Gregg, K., Baydoun, A., Rezai, A., Langmack, C., Henke, L., Spratt, D. E., Kashani, R., & Price, A. (2025). Characterization of a prototype rapid kilovoltage x‐ray image guidance system designed for a linear accelerator radiation therapy unit. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70060
Vancouver
Arsenault T, Arrue J, Gregg K, Baydoun A, Rezai A, Langmack C, et al. Characterization of a prototype rapid kilovoltage x‐ray image guidance system designed for a linear accelerator radiation therapy unit. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70060
BibTeX
@article{Arsenault2025,
author = {Arsenault, Theodore and Arrue, Jonathan and Gregg, Kenneth and Baydoun, Atallah and Rezai, Atefeh and Langmack, Christian and Henke, Lauren and Spratt, Daniel E and Kashani, Rojano and Price, Alex},
title = {Characterization of a prototype rapid kilovoltage x‐ray image guidance system designed for a linear accelerator radiation therapy unit},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70060}
}
EUVIMED