Feasibility of Ethos adaptive treatments of lung tumors and associated quality assurance

Sonja Wegener iD, Stefan Weick, Robert Schindhelm, Jörg Tamihardja iD, Otto A Sauer, Gary Razinskas iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14311 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Sonja Wegener und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiotherapy and Radiation Oncology, University of Wurzburg, Wurzburg, Germany. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Lung Neoplasms, Cone-Beam Computed Tomography, Radiotherapy Dosage. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Würzburg, Julius-Maximilians-Universität Würzburg.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Motivation Online adaptive radiotherapy with Ethos is based on the anatomy determined from daily cone beam computed tomography (CBCT) images. Dose optimization and computation are performed on the density map of a synthetic CT (sCT), a deformable registration of the initial planning CT (pCT) onto the current CBCT. Large density changes as present in the lung region are challenging the system.

Methods

Treatment plans for Ethos were created and delivered for 1, 2, and 3 cm diameter lung lesions in an anthropomorphic phantom, combining different insets in the pCT and during adaptive and non‐adaptive treatment sessions. Primary and secondary dose calculations as well as back‐projected dose from portal images were evaluated.

Results

Density changes due to changed insets were not considered in the sCTs. This resulted in errors in the dose; for example, ‐15.9% of the mean dose for a plan when changing from a 3 cm inset in the pCT to 1 cm at the time of treatment. Secondary dose calculation is based on the sCT and could therefore not reveal these dose errors. However, dose calculation on the CBCT, either as a recalculation in the treatment planning system or as pre‐treatment quality assurance (QA) before the treatment, indicated the differences. EPID in‐vivo QA also reported discrepancies between calculated and delivered dose distributions.

Conclusions

An incorrect density distribution in the sCT has an impact on the dose calculation accuracy in the adaptive treatment workflow with the Ethos system. Additional quality checks of the sCT can detect such errors.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedLung NeoplasmsCone-Beam Computed TomographyRadiotherapy DosageQuality Assurance, Health CarePhantoms, ImagingFeasibility StudiesRadiotherapy, Intensity-ModulatedHumansImage Processing, Computer-AssistedRadiotherapy, Image-GuidedOrgans at RiskAlgorithms

Stichwörter der Autor:innen

EPID dosimetryadaptive radiotherapylungsecondary dose calculationsynthetic CT

Autor:innen und Institutionen

  • Department of Radiotherapy and Radiation Oncology, University of Wurzburg, Wurzburg, Germany.

ORCID-Profile: Sonja Wegener, Jörg Tamihardja, Gary Razinskas

Förderung

  • University of Würzburg
  • Julius-Maximilians-Universität Würzburg

Zitieren

APA

Wegener, S., Weick, S., Schindhelm, R., Tamihardja, J., Sauer, O. A., & Razinskas, G. (2024). Feasibility of Ethos adaptive treatments of lung tumors and associated quality assurance. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14311

Vancouver

Wegener S, Weick S, Schindhelm R, Tamihardja J, Sauer OA, Razinskas G. Feasibility of Ethos adaptive treatments of lung tumors and associated quality assurance. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14311

BibTeX
@article{Wegener2024,
  author = {Wegener, Sonja and Weick, Stefan and Schindhelm, Robert and Tamihardja, Jörg and Sauer, Otto A and Razinskas, Gary},
  title = {Feasibility of Ethos adaptive treatments of lung tumors and associated quality assurance},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14311}
}
Kostenlos starten →