EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Commissioning a clinical linac for preclinical research of heterogeneous intratumor irradiation with end‐to‐end tests

Shuting Wang, MengYao Li, Zhenkai Li, Xiaoying Fan, Yong Yin, Tianyuan Dai iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70402 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Shuting Wang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Graduate, Shandong First Medical University, Shandong Academy of Medical Sciences, Jinan, China. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Particle Accelerators, Phantoms, Imaging, Neoplasms. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Shandong Provincial Natural Science Foundation, Taishan Scholars Project of Shandong Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Although it is not clear how the heterogeneous intratumor irradiation pattern affects the biological effect, investigations of the biological effect dependence on the dose and size of the peak region are essential in the current preclinical research.

Purpose

This work develops and validates a cone‐equipped clinical linac platform for heterogeneous intratumor irradiation (HII) of small animals, making translationally relevant preclinical radiobiology studies accessible to more clinical centers.

Methods

Output ratios (ORs) for cones (from 4 mm to 17.5 mm diameters) were measured using Gafchromic EBT3 films on a Varian TrueBeam linac (6 MV FFF beam). Tissue maximum ratios (TMR) and off‐axis ratios (OAR) were incorporated into the Eclipse treatment planning system (TPS) to configure the beam model. End‐to‐end test was performed using both a mouse phantom and tumor‐bearing mice, with dose distributions verified via film dosimetry and γ analysis (3%/3 mm and 3%/2 mm criteria).

Results

The system achieved millimeter‐scale precision, with γ pass rates ≥ 99% (3%/2 mm). Dose distributions confirmed the feasibility of delivering heterogeneous doses (e.g., 14 Gy via 4 mm cone + 2 Gy via 15 mm cone) while sparing normal tissues.

Conclusion

This study established a robust preclinical HII platform using clinical linac infrastructure, enabling precise dose delivery for investigating dose‐response relationships and supporting future translational research.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedParticle AcceleratorsPhantoms, ImagingNeoplasmsRadiotherapy, Intensity-ModulatedOrgans at RiskAnimalsRadiotherapy DosageMiceHumansFilm Dosimetry

Stichwörter der Autor:innen

Heterogeneous intratumor irradiation (HII)clinical linacconical collimatorsend‐to‐end testsmall animal irradiation

Autor:innen und Institutionen

  • Department of Graduate, Shandong First Medical University, Shandong Academy of Medical Sciences, Jinan, China.
  • Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, China.
  • Department of Oncology, Afliated Hospital of Southwest Medical University, Luzhou, Sichuan, China.

ORCID-Profile: Tianyuan Dai

Förderung

  • National Natural Science Foundation of China
  • Shandong Provincial Natural Science Foundation
  • Taishan Scholars Project of Shandong Province
  • Xinjiang Uygur Autonomous Region
  • Shandong Province Higher Education Institutions Youth Innovation Team Development Plan
  • Young Talent of Lifting engineering for Science and Technology in Shandong, China
  • Tumor precision radiotherapy peak plan
  • Natural Science Foundation of Shandong Province

Zitieren

APA

Wang, S., Li, M., Li, Z., Fan, X., Yin, Y., & Dai, T. (2025). Commissioning a clinical linac for preclinical research of heterogeneous intratumor irradiation with end‐to‐end tests. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70402

Vancouver

Wang S, Li M, Li Z, Fan X, Yin Y, Dai T. Commissioning a clinical linac for preclinical research of heterogeneous intratumor irradiation with end‐to‐end tests. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70402

BibTeX
@article{Wang2025,
  author = {Wang, Shuting and Li, MengYao and Li, Zhenkai and Fan, Xiaoying and Yin, Yong and Dai, Tianyuan},
  title = {Commissioning a clinical linac for preclinical research of heterogeneous intratumor irradiation with end‐to‐end tests},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70402}
}
Kostenlos starten →