EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Reference dosimetry data and machine performance comparison for tomotherapy delivery systems audited by IROC

Paige A Taylor iD, Hunter Mehrens, Mallory C Glenn, Fre'Etta Brooks, Hayden Scott iD, Lian Duan iD, Taylor Meyers, Diana Carrasco Rojas, Michael Yang, Paola Alvarez iD, Jessica Lowenstein, Andrea Molineu und 1 weitere

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70390 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Paige A Taylor und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, The University of MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging, Radiometry. Die Arbeit nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The aim of this work was to compare the imaging and radiation oncology core (IROC) on‐site TomoTherapy dosimetry measurement results among institutions and generate a dataset of basic dosimetric properties for TomoTherapy units. Independent ion chamber measurements for nine TomoTherapy units were acquired by IROC physicists between 2008 and 2023. Measurements included percent depth dose (PDD) in a water tank and off axis factors (OAF) in a solid Polystyrene phantom. The independent measurements collected during the on‐site audits for each TomoTherapy system were compared to corresponding TomoTherapy TPS calculations from the institution to assess agreement. Numeric measurement values and comparative ratios are compiled and tabulated according to measurement type (i.e., PDD and OAF). The distribution of the measurements was evaluated, and the mean, standard deviation of the mean, and median values are reported. These data provide an independent guide describing the inherent performance characteristics of TomoTherapy units. Such data may serve as a secondary check for users wishing to verify their beam model commissioning and ongoing quality assurance efforts.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedPhantoms, ImagingRadiometryQuality Assurance, Health CareNeoplasmsHumansRadiotherapy Dosage

Stichwörter der Autor:innen

TomoTherapyauditquality assurance

Autor:innen und Institutionen

  • Department of Radiation Physics, The University of MD Anderson Cancer Center, Houston, Texas, USA.
  • Medical Physics Graduate Program, The University of Texas MD Anderson Cancer Center Graduate School of Biomedical Sciences, Houston, Texas, USA.
  • University of Washington Department of Radiation Oncology, Seattle, Washington, USA.

ORCID-Profile: Paige A Taylor, Hayden Scott, Lian Duan, Paola Alvarez

Förderung

  • National Institute of Health

Zitieren

APA

Taylor, P. A., Mehrens, H., Glenn, M. C., Brooks, F., Scott, H., Duan, L., Meyers, T., Rojas, D. C., Yang, M., Alvarez, P., Lowenstein, J., Molineu, A., & Kry, S. F. (2025). Reference dosimetry data and machine performance comparison for tomotherapy delivery systems audited by IROC. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70390

Vancouver

Taylor PA, Mehrens H, Glenn MC, Brooks F, Scott H, Duan L, et al. Reference dosimetry data and machine performance comparison for tomotherapy delivery systems audited by IROC. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70390

BibTeX
@article{Taylor2025,
  author = {Taylor, Paige A and Mehrens, Hunter and Glenn, Mallory C and Brooks, Fre'Etta and Scott, Hayden and Duan, Lian and Meyers, Taylor and Rojas, Diana Carrasco and Yang, Michael and Alvarez, Paola and Lowenstein, Jessica and Molineu, Andrea and Kry, Stephen F},
  title = {Reference dosimetry data and machine performance comparison for tomotherapy delivery systems audited by IROC},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70390}
}
Kostenlos starten →