Multi‐institutional study on the commissioning and clinical implementation of the TrueBeam enhanced leaf model in the Eclipse treatment planning system
Journal of Applied Clinical Medical Physics · 2025-12 · 26 (12)
DOI: 10.1002/acm2.70360 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 25.11.2025. Verfasst von Ryohei Miyasaka und 15 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiation Oncology, Chiba Cancer Center, Chiba, Chiba, Japan. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Particle Accelerators, Prostatic Neoplasms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Varian Medical Systems.
Abstract (Originalsprache)
Purpose
Recently, the Enhanced Leaf Modeling (ELM) was developed to improve the multileaf collimator (MLC) model in the Eclipse treatment planning system (TPS). This multi‐institutional study evaluated the dose calculation accuracy and clinical feasibility of the ELM for TrueBeam linear accelerators (linacs) equipped with Millennium 120 and High‐Definition 120 MLCs. In addition, to facilitate broader clinical adoption, the feasibility of using machine‐averaged ELMs (ELM Ave ) for both MLC types was evaluated.
Methods
Twelve TrueBeam linacs from nine institutions were included. Machine‐specific ELM (ELM MS ) parameters, including Leaf Transmission (LT) and the Leaf Gap (LG), were derived from vendor MLC sequences. Three volumetric modulated arc therapy (VMAT) plans for prostate cancer, head and neck cancer, and spinal metastasis were calculated using Acuros XB with each ELM MS and verified using an ionization chamber and either Delta4 or ArcCHECK systems, based on relative dose differences and gamma pass rates (3%/2 mm). Verification criteria were set as tolerance and action levels of ≤2% and ≤3% for relative dose difference and ≥95% and ≥90% for gamma pass rate. To further improve the ELM dose calculation accuracy for VMAT cases, the LT and LG parameters were systematically adjusted based on the verification results and then averaged across all linacs to obtain the ELM Ave, which was subsequently verified using the same procedure.
Results
For the ELM MS dose calculations, the relative dose differences across the three VMAT cases averaged −1.13%, with a two‐standard deviation (2 SD) of 1.04%. The mean gamma pass rate for all dose profiles was 99.4% (2 SD = 2.6%). The ELM MS achieved clinically acceptable accuracy, with 96.3% of the verification results within clinical tolerance and no cases exceeding the action level. After parameter tuning, the ELM Ave calculations exhibited dose differences within ±2% and gamma pass rates above 96% across all verification cases investigated in this study.
Conclusions
The ELM MS demonstrated reliable VMAT dose calculation accuracy across multiple institutions and MLC types. Tuning the LT and LG parameters can further improve ELM performance in various VMAT plans. The tuned ELM Ave satisfied all clinical tolerance criteria in all verification cases and may serve as a practical approach for efficient commissioning and objective evaluation of ELM parameters across institutions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Radiation Oncology, Chiba Cancer Center, Chiba, Chiba, Japan.
- Department of Radiation Oncology, Saitama Medical University International Medical Center, Hidaka, Saitama, Japan.
- Department of Radiation Oncology, Tokyo Metropolitan Cancer and Infectious Diseases Center Komagome Hospital, Bunkyo-ku, Tokyo, Japan.
- Department of Radiation Oncology, St. Luke's International Hospital, Chuo-ku, Tokyo, Japan.
- Department of Radiation Oncology, Saitama Medical Center, Saitama Medical University, Kawagoe, Saitama, Japan.
- Department of Radiological Technology, Tokyo Saiseikai Central Hospital, Minato-ku, Tokyo, Japan.
ORCID-Profile: Ryohei Yamauchi, Norifumi Mizuno, Kaito Sakai
Förderung
- Varian Medical Systems
Zitieren
APA
Miyasaka, R., Hirai, R., Suda, Y., Yamauchi, R., Mizuno, N., Suzuki, Y., Igari, M., Ohta, T., Nakayama, H., Arai, Y., Sakai, K., Hatakeyama, K., Miyashita, H., Kurooka, M., Kawachi, T., & Hara, R. (2025). Multi‐institutional study on the commissioning and clinical implementation of the TrueBeam enhanced leaf model in the Eclipse treatment planning system. Journal of Applied Clinical Medical Physics, 26(12). https://doi.org/10.1002/acm2.70360
Vancouver
Miyasaka R, Hirai R, Suda Y, Yamauchi R, Mizuno N, Suzuki Y, et al. Multi‐institutional study on the commissioning and clinical implementation of the TrueBeam enhanced leaf model in the Eclipse treatment planning system. Journal of Applied Clinical Medical Physics. 2025;26(12). doi:10.1002/acm2.70360
BibTeX
@article{Miyasaka2025,
author = {Miyasaka, Ryohei and Hirai, Ryuta and Suda, Yuhi and Yamauchi, Ryohei and Mizuno, Norifumi and Suzuki, Yuya and Igari, Mitsunobu and Ohta, Tomohiro and Nakayama, Hiroki and Arai, Yu and Sakai, Kaito and Hatakeyama, Kazushi and Miyashita, Hisayuki and Kurooka, Masahiko and Kawachi, Toru and Hara, Ryusuke},
title = {Multi‐institutional study on the commissioning and clinical implementation of the TrueBeam enhanced leaf model in the Eclipse treatment planning system},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
volume = {26},
number = {12},
doi = {10.1002/acm2.70360}
}
EUVIMED