Evaluation of multileaf collimator driving accuracy in helical rotational irradiation system: Quantitative analysis of leaf open time during treatment

Hirofumi Honda, Motoharu Sasaki, Masahide Tominaga, Kenji Omoto, Teruhito Kido

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70186 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Hirofumi Honda und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Ehime University Hospital, Toon, Ehime, Japan. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Prostatic Neoplasms, Head and Neck Neoplasms. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The Radixact treatment system is equipped with a delivery analysis feature. This feature enables dose reconstruction using the patient's treatment‐planning computed tomography scans and allows verification of the multileaf collimator (MLC) performance before and during treatment. In the Radixact system, the opening time of the MLC leaves is determined based on the treatment plan.

Purpose

This study aimed to evaluate MLC driving accuracy by assessing the MLC leaf open time (LOT) during treatment.

Methods

Using Delivery Analysis version 2.3, we compared the treatment plan LOT with the LOT measured during treatment to determine the average and one standard deviation (%) of the LOT attainment rate. The analysis included comparisons of treated sites across 39 cases: nine prostate, eight pelvic, seven head, six chest, five head and neck (H&N), and four stereotactic body radiation therapy (SBRT) treatment plans.

Results

The average and one standard deviation (%) of the LOT attainment rate for all patients on treatment was 94.56 ± 2.37. The values of each site were as follows: prostate, 95.93 ± 0.68; pelvis, 93.37 ± 2.16; head, 95.05 ± 1.99; chest, 97.61 ± 0.78; H&N, 92.44 ± 1.32; and SBRT, 98.39 ± 0.57. The treatment plans with the lowest attainment rates for each site were as follows: prostate, 95.19 ± 0.39; pelvis, 90.59 ± 0.16; head, 92.20 ± 0.15; chest, 95.76 ± 0.04; H&N, 90.55 ± 0.30; and SBRT, 97.32 ± 0.07. The plans with the largest one standard deviation (%) per site were as follows: prostate, ± 0.97; pelvis, ± 0.26; head, ± 0.57; chest, ± 0.23; H&N, ± 0.30; and SBRT, ± 0.07.

Conclusions

We proposed a simple method for quantitatively analyzing the LOT of an MLC. The average LOT attainment rate and its standard deviation varied by treatment site. Since the standard deviation differed by plan, the LOT attainment rate during treatment should be carefully monitored.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedProstatic NeoplasmsHead and Neck NeoplasmsRadiosurgeryParticle AcceleratorsNeoplasmsHumansMaleRadiotherapy DosageOrgans at RiskThoracic NeoplasmsPelvic Neoplasms

Stichwörter der Autor:innen

MLC leaf open timeRadixact systemdelivery analysismulti‐leaf collimator

Autor:innen und Institutionen

  • Department of Radiological Technology, Ehime University Hospital, Toon, Ehime, Japan.
  • Graduate School of Advanced Technology and Science, Tokushima University, Tokushima, Japan.
  • Department of Radiology, Ehime University Graduate School of Medicine, Toon, Ehime, Japan.

Zitieren

APA

Honda, H., Sasaki, M., Tominaga, M., Omoto, K., & Kido, T. (2025). Evaluation of multileaf collimator driving accuracy in helical rotational irradiation system: Quantitative analysis of leaf open time during treatment. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70186

Vancouver

Honda H, Sasaki M, Tominaga M, Omoto K, Kido T. Evaluation of multileaf collimator driving accuracy in helical rotational irradiation system: Quantitative analysis of leaf open time during treatment. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70186

BibTeX
@article{Honda2025,
  author = {Honda, Hirofumi and Sasaki, Motoharu and Tominaga, Masahide and Omoto, Kenji and Kido, Teruhito},
  title = {Evaluation of multileaf collimator driving accuracy in helical rotational irradiation system: Quantitative analysis of leaf open time during treatment},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70186}
}
Kostenlos starten →