Effective spot size parameters for Acuros dose calculation algorithm using enhanced leaf modelling: Estimation based on small rectangular MLC fields

Antonella Fogliata iD, Antonella Stravato iD, Marco Pelizzoli, Francesco La Fauci iD, Pasqualina Gallo iD, Andrea Bresolin, Luca Cozzi, Giacomo Reggiori

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70315 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Antonella Fogliata und 7 weiteren Autor:innen. Beteiligte Einrichtung: Radiotherapy and Radiosurgery Department, Humanitas Research Hospital IRCCS, Milan-Rozzano, Italy. Hauptthemen laut MEDLINE: Algorithms, Radiotherapy Planning, Computer-Assisted, Particle Accelerators, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Purpose Small field dosimetry plays a critical role in stereotactic radiotherapy, particularly for elongated fields shaped by multileaf collimators (MLCs). This study aims to optimize the Effective Spot Size (ESS) parameters in the Acuros dose calculation algorithm (Eclipse v18.0, with the enhanced leaf modelling ELM for MLC modelling) for accurate dose calculations of small, rectangular MLC‐shaped fields. Measurements were based on the TRS‐483 protocol and a recently proposed equivalent square field size determination method.

Methods

Field output factors (FOFs) were measured using a Varian TrueBeam linear accelerator with HD‐MLC for 6 MV and 10 MV beams (both flattened and unflattened). Symmetric square and rectangular fields (0.5–4 cm sides) were shaped by the MLC, with jaws fixed at 4.4 × 4.4 cm 2 to minimize scatter variation. Measurements were performed at isocenter, 10 cm depth using three detectors: microDiamond, PinPoint3D, and DiodeE. The microDiamond detector results were used for ESS optimization due to its complete output correction factor availability. FOFs were calculated in Eclipse v18.0 using Acuros with varying ESS values (0–1.5 mm). The optimal ESS was determined by minimizing FOF differences between calculated and measured values for extreme elongated field sizes in the studied range (0.5 × 4 cm 2 and 4 × 0.5 cm 2 ).

Results

The optimized ESS parameters minimized FOF discrepancies between calculated and measured doses for elongated fields, improving calculation accuracy compared to vendor‐recommended settings (ESSx = 0.5 mm, ESSy = 0.7 mm). Profiles and penumbra analyses supported these findings.

Conclusions

ELM in Eclipse v.18 improves MLC modelling by representing rounded leaf ends. Only ESS parameters remain user‐tunable. A method to determine the ESS for the Acuros algorithm is presented, resulting in accurate algorithm configuration for elongated small fields shaped by an HD‐MLC. Results are presented using 6 MV and 10 MV beams, both flattened and unflattened for an HD‐MLC Varian. ESS affects output factors and penumbra, with dose calculations matching measurements within 0.2% ± 0.4% and a maximum difference of 1.1%.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsRadiotherapy Planning, Computer-AssistedParticle AcceleratorsRadiotherapy, Intensity-ModulatedPhantoms, ImagingNeoplasmsRadiosurgeryRadiotherapy DosageHumansRadiometry

Stichwörter der Autor:innen

Acuroseffective spot sizeoutput factorsrectangular small fieldssmall fields

Autor:innen und Institutionen

  • Radiotherapy and Radiosurgery Department, Humanitas Research Hospital IRCCS, Milan-Rozzano, Italy.
  • Medical Physics Department, A.O. San Giovanni-Addolorata, Roma, Italy.
  • UniCamillus, International Medical University in Rome, Roma, Italy.

ORCID-Profile: Antonella Fogliata, Antonella Stravato, Francesco La Fauci, Pasqualina Gallo

Zitieren

APA

Fogliata, A., Stravato, A., Pelizzoli, M., La Fauci, F., Gallo, P., Bresolin, A., Cozzi, L., & Reggiori, G. (2025). Effective spot size parameters for Acuros dose calculation algorithm using enhanced leaf modelling: Estimation based on small rectangular MLC fields. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70315

Vancouver

Fogliata A, Stravato A, Pelizzoli M, La Fauci F, Gallo P, Bresolin A, et al. Effective spot size parameters for Acuros dose calculation algorithm using enhanced leaf modelling: Estimation based on small rectangular MLC fields. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70315

BibTeX
@article{Fogliata2025,
  author = {Fogliata, Antonella and Stravato, Antonella and Pelizzoli, Marco and La Fauci, Francesco and Gallo, Pasqualina and Bresolin, Andrea and Cozzi, Luca and Reggiori, Giacomo},
  title = {Effective spot size parameters for Acuros dose calculation algorithm using enhanced leaf modelling: Estimation based on small rectangular MLC fields},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70315}
}
Kostenlos starten →