New gantry angle‐dependent beam control optimization with Elekta linear accelerator for VMAT delivery
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.14598 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Adriaan Abraham van Appeldoorn und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiotherapy, University Medical Center Utrecht, Utrecht, Netherlands. Hauptthemen laut MEDLINE: Particle Accelerators, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Algorithms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 5 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
This paper describes a method to improve gantry‐dependent beam steering for Elekta traveling wave linear accelerators by applying the measured and filtered beam servo corrections to the existing lookup table (LUT). Beam steering has a direct influence on the treatment accuracy by affecting the beam symmetry and position. The presented method provides an improved LUT with respect to the default Elekta method to reduce treatment delivery interruptions. These interruptions are known to contribute to unwanted intrafraction motion and longer treatment times.
Methods
Compared to the default method of the manufacturer, this new method takes both clockwise and counterclockwise rotation to compensate for magnetic hysteresis as well as previous configuration and noise filtering into account. The improved method to determine the lookup table uses service graphing information from the linac without the need for additional symmetry information. The clinical configuration of the flattened beam energies remains untouched during the data record.
Results
This method results in a configuration where the gantry‐dependent steering is optimized over the full arc with optimal balance in the hysteresis and minimizing the effect of errors in the steering values. This method is a less error‐prone process compared to the methodology described in previous research, still achieving a reduction of interruption of about 60 percent compared to the Elekta method.
Conclusion
This study shows a simplified way to optimize linac stability with improved LUT. The optimized LUT results in a lower number of interruptions, preventing downtime, and a lower risk of intrafraction motion due to longer treatment time.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiotherapy, University Medical Center Utrecht, Utrecht, Netherlands.
ORCID-Profile: Adriaan Abraham van Appeldoorn
Zitieren
APA
van Appeldoorn, A. A., Kok, J. G. M., & Wolthaus, J. W. H. (2025). New gantry angle‐dependent beam control optimization with Elekta linear accelerator for VMAT delivery. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14598
Vancouver
van Appeldoorn AA, Kok JGM, Wolthaus JWH. New gantry angle‐dependent beam control optimization with Elekta linear accelerator for VMAT delivery. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14598
BibTeX
@article{vanAppeldoorn2025,
author = {van Appeldoorn, Adriaan Abraham and Kok, Johannes Gerardus Maria and Wolthaus, Jochem Willem Heiko},
title = {New gantry angle‐dependent beam control optimization with Elekta linear accelerator for VMAT delivery},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.14598}
}
EUVIMED